Cold atom clocks have made remarkable progresses in the last two decades and played critical roles in precision measurements. Primary Cs fountain frequency standards have achieved a total uncertainty of a few parts in 1016, and the best optical clock has reached a type B uncertainty below 10−18. Besides applications in the metrology, navigation, etc., ultra-stable and ultra-accurate atomic clocks have also become powerful tools in the basic scientific investigations. In this paper, we focus on the recent developments in the high-performance cold atomic clocks which can be used as frequency standards to calibrate atomic time scales. The basic principles, performances, and limitations of fountain clocks and optical clocks based on signal trapped ion or neutral atoms are summarized. Their applications in metrology and other areas are briefly introduced.
Many efforts have been taken in recent years to push atomic gravimeters toward practical applications. We demonstrate an atomic gravimeter named NIM-AGRb2 that is transportable and suitable for high-precision gravity measurements. Constraint-structured active vibration isolation (CS-AVI) is used to reduce the ground vibration noise. The constraint structure in CS-AVI ensures that the isolation platform only has vertical translation, with all other degrees of freedom (DoFs) being constrained. Therefore, the stability of active vibration isolation is enhanced. With the implementation of CS-AVI, the sensitivity of NIM-AGRb2 reached as low as 20.5 μGal/Hz1/2. The short-term sensitivity could be further reduced to 10.8 μGal/Hz1/2 in a seismologic observation station. Moreover, we evaluated the system noise of the gravimeter, and the results were consistent with our observations.
A new Rb fountain clock was built at the National Institute of Metrology with the aim to realize robustness and excellent long-term instability for time keeping. While some basic designs have been adopted from our Cs fountain clocks, new features have been included for improved performance. A double-metal interrogation microwave cavity with a thermal expansion self-compensating mechanism was used to reduce the clock sensitivity to temperature fluctuations for the first time, to our knowledge. Moreover, a compact optical system was developed to ensure robustness. These features dramatically increased the ambient temperature range. The developed Rb fountain clock achieved a typical fractional frequency instability of within 10 days of continuous operation without any intervention, demonstrating the potential for time keeping.
23 November 2022 in IEEE Transactions on Microwave Theory and Techniques
A typical Ramsey cavity, commonly used in atomic fountain frequency standards (AFFSs) for microwave interrogation, is sensitive to ambient temperature variation. Then, it becomes one of the main limiting factors for the AFFSs operating in a narrow operating-temperature range and a temperature-well-controlled environment up to now. In this article, a new design of bimetal temperature-compensated Ramsey cavity (BTCRC) based on the thermal expansion self-compensating mechanism is presented and experimentally demonstrated. This cavity is a typical TE $_{011}$ -mode cylindrical microwave resonator with a titanium tube and two oxygen-free copper (OFC) caps. The resonance frequency (RF) of the cavity can be immune to temperature variations if the dimensions of the tube and caps are properly designed. To validate this self-compensating mechanism, a BTCRC resonating at the Rb clock frequency with a $Q$ -factor of about 17 000 is constructed. And its RF thermal coefficient is measured to be 0.73 kHz/ $^{\circ}$ C, which is about 150 times superior to $-$ 115.9 kHz/ $^{\circ}$ C of a typical OFC Rb cavity in the temperature range of 22.48 $^{\circ}$ C–37.48 $^{\circ}$ C. During the operation of AFFSs, this BTCRC ensures that the microwave field amplitude inside the cavity, which dictates the transition probability and thus the contrast of the Ramsey fringes, will be stable in time. In addition, the potential temperature variation-induced frequency shifts due to cavity pulling are significantly minimized.
A new caesium (Cs) fountain clock NIM6 has been developed at the National Institute of Metrology (NIM) in China, for which a comprehensive uncertainty evaluation is presented. A three-dimensional magneto-optical trap loading optical molasses is employed to obtain more cold atoms rapidly and efficiently with a tunable, uniform density distribution. A heat pipe surrounding the flight tube maintains a consistent and stable temperature within the interrogation region. Additionally, a Ramsey cavity with four azimuthally distribution feeds is utilized to mitigate distributed cavity phase shifts. The Cs fountain clock NIM6 achieves a short-term fractional frequency instability of 1.0 × 10 −13τ−1/2 at high atomic density, and a typical overall fractional type-B uncertainty is estimated to be 2.2 × 10 −16 . Comparisons of frequency between the Cs fountain NIM6 and other Cs fountain primary frequency standards through Coordinated Universal Time have demonstrated an agreement within the stated uncertainties.
1 June 2016 in Journal of Physics: Conference Series
The cesium fountain primary frequency standard NIM5 started to operate since 2008 and started to report to BIPM since 2014. The major constrains of NIM5 is a relatively large background signal at the detection and microwave leakages due to the Ramsey cavity. A new fountain clock NIM6 is under construction. Besides some improvements on the vacuum system, a new Ramsey cavity and a microwave synthesizer are made to reduce the Type B uncertainty. Another feature of NIM6 is collecting atoms from a MOT loading optical molasses to get more atoms with a more uniform density distribution. With a new frequency synthesizer based on cryogenic sapphire oscillator (CSO), NIM6 is aiming to reach the quantum projection noise, thus leading to a reduced Type A uncertainty compared with NIM5.