This paper describes the development of an integrated system using a dry film resistant (DFR) microfluidic channel consisting of pulsed field dielectrophoretic field-flow-fractionation (DEP-FFF) separation and optical detection. The prototype chip employs the pulse DEP-FFF concept to separate the cells (Escherichia coli and Saccharomyces cerevisiae) from a continuous flow, and the rate of release of the cells was measured. The separation experiments were conducted by changing the pulsing time over a pulsing time range of 2–24 s and a flow rate range of 1.2–9.6 L min . The frequency and voltage were set to a constant value of 1 Hz and 14 pk-pk, respectively. After cell sorting, the particles pass the optical fibre, and the incident light is scattered (or absorbed), thus, reducing the intensity of the transmitted light. The change in light level is measured by a spectrophotometer and recorded as an absorbance spectrum. The results revealed that, generally, the flow rate and pulsing time influenced the separation of E. coli and S. cerevisiae. It was found that E. coli had the highest rate of release, followed by S. cerevisiae. In this investigation, the developed integrated chip-in-a lab has enabled two microorganisms of different cell dielectric properties and particle size to be separated and subsequently detected using unique optical properties. Optimum separation between these two microorganisms could be obtained using a longer pulsing time of 12 s and a faster flow rate of 9.6 L min at a constant frequency, voltage, and a low conductivity.
This paper describes the development of a lab-on-chip (LOC) device that can perform reliable online detection in continuous-flow systems for microorganisms. The objective of this work was to examine the performance of a fibre optic detection system integrated into a LOC device. The microfluidic system was fabricated using dry film resist (DFR), integrated with multimode fibre pigtails in the LOC. Subsequently, the performance of the fibre optic detection was evaluated by its absorbance spectra, detection limit, repeatability and reproducibility, and response time. The analysis was carried out using a constant flow rate for three different types of microorganisms which are Escherichia coli, Saccharomyces cerevisiae, and Aeromonas hydrophila. Under the experimental conditions used in this study, the detection limit of cells/mL for both A. hydrophila and E. coli, while a detection limit of cells/mL for S. cerevisiae cells were measured. The results also revealed that the device showed good repeatability with standard deviations less than 0.2 for A. hydrophila and E. coli, while standard deviations for S. cerevisiae were larger than 1.0. The response times for A. hydrophila, E. coli, and S. cerevisiae were 104 s, 122 s, and 78 s, respectively, although significant errors were recorded for all three species for reproducibility experiment. It was found that the device showed generally good sensitivity, with the highest sensitivity towards S. cerevisiae. These findings suggest that an integrated LOC device, with embedded multimode fibre pigtails, can be a reliable instrument for microorganism detection.