Ventricular assist devices (VADs) which employ a maglev motor have significant advantages of high mechanical durability and better blood compatibility by eliminating mechanical contact for impeller suspension. In the last decade, research interest in maglev VADs for small pediatric patients is highly increased. However, drastic miniaturization of maglev motors has technical difficulties in deterioration of suspension stability and energy efficiency. Our group has developed an ultra-compact axial gap double stator self-bearing motor utilizing 5-degrees of freedom control for pediatric VADs to overcome the above technical drawbacks. In this article, feasibility of the proposed self-bearing concept for stable impeller suspension was verified, as well as enhancement of energy efficiency of the developed self-bearing motor was investigated by modifying motor core material and winding design. In pump operation with the modified motor, noncontact impeller suspension with energy input of 3.2–5.2 W over a range of pediatric circulatory support (4000–5000 r/mim) was demonstrated.
Mechanical circulatory support (MCS) therapy plays a significant role in an alternative therapy of heart transplants for pediatric heart disease patients. However, continuous flow rotary MCS devices for pediatric patients are still undergoing development, and have not been clinically available technology. Technical difficulties, such as high durability, better blood compatibility and miniature device size, prevent the pediatric MCS devices development. In this study, a double stator axial gap maglev motor for pediatric MCS device has been developed. The maglev motor has two identical motor stators, and a levitated rotor impeller which is aligned between the stators. The levitated rotor impeller is fully suspended with 5-degrees of freedom (5-DOF) active control. A double stator mechanism enhances motor torque production. A miniaturized maglev motor was designed and developed based on FEM magnetic field analysis for use in implantable ventricular assist devices (VADs). The diameter and height of the developed maglev motor are 22 mm and 33 mm. This paper is an initial report on the magnetic levitation and rotation performance of the miniaturized maglev motor. The levitated rotor impeller was magnetically levitated and rotated with the 5-DOF active control. The oscillation amplitudes (x, y and z) and inclination angles (θx and θy) of the levitated rotor impeller were then evaluated in both air and water. The developed maglev motor achieved non-contact rotation up to 1600 rpm in air and 4500 rpm in water, respectively. The oscillation amplitudes and inclination angles were sufficiently suppressed in water due to fluid damping. After these experiments, a magnetic circuit of the maglev motor was modified in order to achieve further stable levitation. The developed maglev motor then indicated potential to achieve the practical use of maglev rotatory pediatric VAD.
29 December 2016 in International Journal of Applied Electromagnetics and Mechanics
An axial gap maglev motor with double stators that can control five-degrees of freedom (5-DOF) of rotor postures has been newly developed for pediatric ventricular assist devices (VADs). The motor has two stators that have an identical structure and
11 May 2020 in The International Journal of Artificial Organs
Implantable ventricular assist devices are a type of mechanical circulatory support for assisting the pumping of the heart. Accurate estimation of the flow rate through such devices is critical to ensure effective performance. A novel method for estimating the flow rate using the passively stabilized position of a magnetically levitated impeller was developed by our group. However, the performance of the method is affected by the gravity vector, which depends on the patient’s posture. In this study, the effects of gravity on the flow estimation method are analyzed, and a compensation method is proposed. The magnetically levitated impeller is axially suspended and radially restricted by passive stability in a centrifugal blood pump developed by our group. The gravity effects were evaluated by analyzing the relationships between the radial position of the magnetically levitated impeller and the flow rate with respect to the gravity direction. Accurate estimation of the flow rate could not be achieved when the direction of gravity with respect to impeller was unknown. A mean absolute error of up to 4.89 L/min was obtained for flow rate measurement range of 0–5 L/min. However, analysis of the equilibrium of forces on the passively stabilized impeller indicated that the effects of gravity on the flow estimation could be compensated by performing additional measurements of the gravity direction with respect to impeller. The method for compensating the effects of gravity on the flow estimation should improve the performance of therapy with the implantable ventricular assist devices.
Research interests of compact magnetically levitated motors have been strongly increased in development of durable and biocompatible mechanical circulatory support (MCS) devices for pediatric heart disease patients. In this study, an ultra-compact axial gap type self-bearing motor with 5-degrees of freedom (DOF) active control for use in pediatric MCS devices has been developed. The motor consists of two identical motor stators and a centrifugal levitated rotor. This paper investigated a design improvement of the magnetic circuit for the self-bearing motor undergoing development in order to diminish energy input by enhancing magnetic suspension and rotation performances. Geometries of the motor were refined based on numerical calculation and three-dimensional (3D) magnetic field analysis. The modified motor can achieve higher suspension force and torque characteristics than that of a previously developed prototype motor. Oscillation of the levitated rotor was significantly suppressed by 5-DOF control over rotating speeds up to 7000 rpm with lower energy input, indicating efficacy of the design refinement of the motor.
1 January 2022 in Journal of the Japan Society of Applied Electromagnetics and Mechanics
An ultra-compact ventricular assist device with an axial gap type self-bearing motor has been developed for use in pediatric circulatory support. The developed self-bearing motor has double stator structure to actively control five degrees of freedom (DOF) of levitated impeller postures and to enhance motor torque capacity with small device size which is 22 mm in outer diameter and 33.6 mm in total height, respectively. This study investigated magnetic suspension stability and energy consumption characteristics during pump operation with 3-DOF (z, θx, θy) control and 5-DOF (x, y, z, θx, θy) control. The axial position and inclination angles were sufficiently stabilized independently with the radial position control. The radial position control can be effective to suppress the radial whirling oscillation at resonance, however power consumption increased by up to 0.4-0.9 W compared to in the case of the motor utilizing 3-DOF control.