Thermal diffusivity of GNPs (graphene nano-platelets) is an important thermo-physical property as it is useful to predict the material behavior in many heat transfer applications. GNP samples were pressed at different loads to obtain different densities, and then thermal diffusivity was measured with the flash method. All samples were coated with a thin layer (~1 µm) of colloidal graphite (Aquadag®) on both sides to reduce reflectance of their surfaces and consequently increase the emissivity. Carrying out measurements on both samples with and without coating, a difference between the two series of measurements was found: This is attributed to a non-negligible transmittance of the uncoated samples due to the porosity of GNPs. Furthermore, assuming a spatial distribution of the light within the samples according to the Lambert-Bougert-Beer law, the extinction coefficient of GNP at different densities has been evaluated processing experimental data with a nonlinear least square regression, (NL-LSF, nonlinear least square fitting), whose model contains the extinction coefficient as unknown. The proposed method represents a further improvement of thermal diffusivity data processing, crucial to calculate the extinction coefficient when data with and without coating are available; or to correct biased thermal diffusivity data when the extinction coefficient is already known. Moreover, reflectance effects have been highlighted comparing asymptotic temperature reached during the tests on coated and uncoated samples at different densities. In fact, the decrease of asymptotic temperature of the uncoated samples gives the percentage of the light reflected and consequently an estimate of the reflectance of the GNP surface.
12 January 2023 in ASME Journal of Heat and Mass Transfer
Four probes for measuring thermal conductivity, thermal diffusivity, and convection heat transfer coefficient were designed, built, and tested. In two of these probes (SP-1 and SP-2), three thermocouples were placed at 25%, 50%, and 75% of the total length of 150 mm, while the third probe (SP-3) has the three thermocouples at 10%, 50%, and 90%. The fourth probe (DP-1) is a dual-probe, to measure thermal diffusivity using the pulse method. First results show good performance of the devices when measuring a reference material (glycerol). In fact, an accuracy of about ±5% was achieved in both thermal conductivity and thermal diffusivity measurements. The size of the probes (length 150 mm, diameter 2 mm) makes them particularly suitable for measuring the thermophysical properties of bulk materials, such as soils, composites, buildings walls, etc. Thanks to the presence of three thermocouples, convection can be studied so that also the boundary layer can also be evaluated.