Human grasping and manipulation control critically depends on tactile feedback. Without this feedback, the ability for fine control of a prosthesis is limited in upper limb amputees. Although various approaches have been investigated in the past, at present there is no commercially available device able to restore tactile feedback in upper limb amputees. Based on the Discrete Event-driven Sensory feedback Control (DESC) policy we present a device able to deliver short-lasting vibrotactile feedback to transradial amputees using commercially available myoelectric hands. The device (DESC-glove) comprises sensorized thimbles to be placed on the prosthesis digits, a battery-powered electronic board, and vibrating units embedded in an arm-cuff being transiently activated when the prosthesis makes and breaks contact with objects. The consequences of using the DESC-glove were evaluated in a longitudinal study. Five transradial amputees were equipped with the device for one month at home. Through a simple test proposed here for the first time-the virtual eggs test-we demonstrate the effectiveness of the device for prosthetic control in daily life conditions. In the future the device could be easily exploited as an add-on to complement myoelectric prostheses or even embedded in prosthetic sockets to enhance their control by upper limb amputees.
9 June 2016 in IEEE Transactions on Neural Systems and Rehabilitation Engineering
The replacement of a missing hand by a prosthesis is one of the most fascinating challenges in rehabilitation engineering. State of art prostheses are curtailed by the physical features of the hand, like poor functionality and excessive weight. Here we present a new multi-grasp hand aimed at overcoming such limitations. The SSSA-MyHand builds around a novel transmission mechanism that implements a semi-independent actuation of the abduction/adduction of the thumb and of the flexion/extension of the index, by means of a single actuator. Thus, with only three electric motors the hand is capable to perform most of the grasps and gestures useful in activities of daily living, akin commercial prostheses with up to six actuators, albeit it is as lightweight as conventional 1-Degrees of Freedom prostheses. The hand integrates position and force sensors and an embedded controller that implements automatic grasps and allows inter-operability with different human-machine interfaces. We present the requirements, the design rationale of the first prototype and the evaluation of its performance. The weight (478 g), force (31 N maximum force at the thumb fingertip) and speed of the hand (closing time: <;370 ms), make this new design an interesting alternative to clinically available multi-grasp prostheses.
15 July 2014 in IEEE Transactions on Neural Systems and Rehabilitation Engineering
Tactile feedback is essential to intuitive control and to promote the sense of self-attribution of a prosthetic limb. Recent findings showed that amputees can be tricked to experience this embodiment, when synchronous and modality-matched stimuli are delivered to biological afferent structures and to an alien rubber hand. Hence, it was suggested to exploit this effect by coupling touch sensors in a prosthesis to an array of haptic tactile stimulators in the prosthetic socket. However, this approach is not clinically viable due to physical limits of current haptic devices. To address this issue we have proposed modality-mismatched stimulation and demonstrated that this promotes self-attribution of an alien hand on normally limbed subjects. In this work we investigated whether similar effects could be induced in transradial amputees with referred phantom sensations in a series of experiments fashioned after the Rubber Hand Illusion using vibrotactile stimulators. Results from three independent measures of embodiment demonstrated that vibrotactile sensory substitution elicits body-ownership of a rubber hand in transradial amputees. These results open up promising possibilities in this field; indeed miniature, safe and inexpensive vibrators could be fitted into commercially available prostheses and sockets to induce the illusion every time the prosthesis manipulates an object.
Objective . Tactile afferents in the human hand provide fundamental information about hand-environment interactions, which is used by the brain to adapt the motor output to the physical properties of the object being manipulated. A hand amputation disrupts both afferent and efferent pathways from/to the hand, completely invalidating the individual’s motor repertoire. Although motor functions may be partially recovered by using a myoelectric prosthesis, providing functionally effective sensory feedback to users of prosthetics is a largely unsolved challenge. While past studies using invasive stimulation suggested that sensory feedback may help in handling fragile objects, none explored the underpinning, relearned, motor coordination during grasping. In this study, we aimed at showing for the first time that intraneural sensory feedback of the grip force (GF) improves the sensorimotor control of a transradial amputee controlling a myoelectric prosthesis. Approach . We performed a longitudinal study testing a single subject (clinical trial registration number NCT02848846). A stacking cups test (CUP) performed over two weeks aimed at measuring the subject’s ability to finely regulate the GF applied with the prosthesis. A pick and lift test (PLT), performed at the end of the study, measured the level of motor coordination, and whether the subject transferred the motor skills learned in the CUP to an alien task. Main results . The results show that intraneural sensory feedback increases the subject’s ability in regulating the GF and allows for improved performance over time. Additionally, the PLT demonstrated that the subject was able to generalize and transfer her manipulation skills to an unknown task and to improve her motor coordination. Significance . Our findings suggest that intraneural sensory feedback holds the potential of restoring functionally effective tactile feedback. This opens up new possibilities to improve the quality of life of amputees using a neural prosthesis.
Replacement of a lost limb by an artificial substitute is not yet ideal. Resolution and coordination of motor control approximating that of a biological limb could dramatically improve the functionality of prosthetic devices, and thus reduce the gap towards a suitable limb replacement. In this study, we investigated the control resolution and coordination exhibited by subjects with transhumeral amputation who were implanted with epimysial electrodes and an osseointegrated interface that provides bidirectional communication in addition to skeletal attachment (e-OPRA Implant System). We assessed control resolution and coordination in the context of routine and delicate grasping using the Pick and Lift and the Virtual Eggs Tests. Performance when utilizing implanted electrodes was compared with the standard-of-care technology for myoelectric prostheses, namely surface electrodes. Results showed that implanted electrodes provide superior controllability over the prosthetic terminal device compared to conventional surface electrodes. Significant improvements were found in the control of the grip force and its reliability during object transfer. However, these improvements failed to increase motor coordination, and surprisingly decreased the temporal correlation between grip and load forces observed with surface electrodes. We found that despite being more functional and reliable, prosthetic control via implanted electrodes still depended highly on visual feedback. Our findings indicate that incidental sensory feedback (visual, auditory, and osseoperceptive in this case) is insufficient for restoring natural grasp behavior in amputees, and support the idea that supplemental tactile sensory feedback is needed to learn and maintain the motor tasks internal model, which could ultimately restore natural grasp behavior in subjects using prosthetic hands.