A portable hybrid brain monitoring system is proposed to perform simultaneous 16-channel electroencephalogram (EEG) and 8-channel functional near-infrared spectroscopy (fNIRS) measurements. Architecture-optimized analog frontend integrated circuits (Texas Instruments ADS1299 and ADS8688A) were used to simultaneously achieve 24-bit EEG resolution and reliable latency-less (<;0.85 μs) bio-optical measurements. Suppression of the noise and crosstalk generated by the digital circuit components and flashing NIR light sources was maximized through linear regulator-based fully isolated circuit design. Gel-less EEG measurements were enabled by using spring-loaded dry electrodes. Several evaluations were carried out by conducting an EEG phantom test and an arterial occlusion experiment. An alpha rhythm detection test (eye-closing task) and a mental arithmetic experiment (cumulative subtraction task) were conducted to determine whether the system is applicable to human subject studies. The evaluation results show that the proposed system is sufficiently capable of detecting microvoltage EEG signals and hemodynamic responses. The results of the studies on human subjects enabled us to verify that the proposed system is able to detect task-related EEG spectral features such as eye-closed event-related synchronization and mental-arithmetic event-related desynchronization in the alpha and beta rhythm ranges. An analysis of the fNIRS measurements with an arithmetic operation task also revealed a decreasing trend in oxyhemoglobin concentration.
Dry contact electrode-based EEG acquisition is one of the easiest ways to obtain neural information from the human brain, providing many advantages such as rapid installation, and enhanced wearability. However, high contact impedance due to insufficient electrical coupling at the electrode-scalp interface still remains a critical issue. In this paper, a two-wired active dry electrode system is proposed by combining finger-shaped spring-loaded probes and active buffer circuits. The shrinkable probes and bootstrap topology-based buffer circuitry provide reliable electrical coupling with an uneven and hairy scalp and effective input impedance conversion along with low input capacitance. Through analysis of the equivalent circuit model, the proposed electrode was carefully designed by employing off-the-shelf discrete components and a low-noise zero-drift amplifier. Several electrical evaluations such as noise spectral density measurements and input capacitance estimation were performed together with simple experiments for alpha rhythm detection. The experimental results showed that the proposed electrode is capable of clear detection for the alpha rhythm activation, with excellent electrical characteristics such as low-noise of 1.131 μVRMS and 32.3% reduction of input capacitance.