Aiming at continuous unobtrusive respiration monitoring, motion robustness is paramount. However, some types of motion can completely hide the respiration information and the detection of these events is required to avoid incorrect rate estimations. Therefore, this work proposes a motion detector optimized to specifically detect severe motion of infants combined with a respiration rate detection strategy based on automatic pixels selection, which proved to be robust to motion of the infants involving head and limbs. A dataset including both thermal and RGB (Red Green Blue) videos was used amounting to a total of 43 h acquired on 17 infants. The method was successfully applied to both RGB and thermal videos and compared to the chest impedance signal. The Mean Absolute Error (MAE) in segments where some motion is present was and breaths/min higher than the MAE in the ideal moments where the infants were still for testing and validation set, respectively. Overall, the average MAE on the testing and validation set are breaths/min and breaths/min, using and of the included video segments (segments containing events such as interventions were excluded based on a manual annotation), respectively. Moreover, we highlight challenges that need to be overcome for continuous camera-based respiration monitoring. The method can be applied to different camera modalities, does not require skin visibility, and is robust to some motion of the infants.
In this paper, we investigate an automated pipeline to estimate respiration signals from videos for premature infants in neonatal intensive care units (NICUs). Two flow estimation methods, namely the conventional optical flow- and deep learning-based flow estimation methods, were employed and compared to estimate pixel motion vectors between adjacent video frames. The respiratory signal is further extracted via motion factorization. The proposed methods were evaluated by comparing our automated extracted respiration signals to that extracted from chest impedance on videos of five premature infants. The overall average cross-correlation coefficients are 0.70 for the optical flow-based method and 0.74 for the deep flow-based method. The average root mean-squared errors are 6.10 and 4.55 for the optical flow- and the deep flow-based methods, respectively. The experimental results are promising for further investigation and clinical application of the video-based respiration monitoring method for infants in NICUs.
Both Respiratory Flow (RF) and Respiratory Motion (RM) are visible in thermal recordings of infants. Monitoring these two signals usually requires landmark detection for the selection of a region of interest. Other approaches combine respiratory signals coming from both RF and RM, obtaining a Mixed Respiratory (MR) signal. The detection and classification of apneas, particularly common in preterm infants with low birth weight, would benefit from monitoring both RF and RM, or MR, signals. Therefore, we propose in this work an automatic RF pixel detector not based on facial/body landmarks. The method is based on the property of RF pixels in thermal videos, which are in areas with a smooth circular gradient. We defined 5 features combined with the use of a bank of Gabor filters that together allow selection of the RF pixels. The algorithm was tested on thermal recordings of 9 infants amounting to a total of 132 min acquired in a neonatal ward. On average the percentage of correctly identified RF pixels was . Obstructive Apneas (OAs) were simulated as a proof of concept to prove the advantage in monitoring the RF signal compared to the MR signal. The sensitivity in the simulated OA detection improved for the RF signal reaching against the of the MR signal. Overall, the method yielded promising results, although the positioning and number of cameras used could be further optimized for optimal RF visibility.
Low-resolution thermal cameras have already been used in the detection of respiratory flow. However, microbolometer technology has a high production cost compared to thermopile arrays. In this work, the feasibility of using a thermopile array to detect respiratory flow has been investigated in multiple settings. To prove the concept, we tested the detector on six healthy subjects. Our method automatically selects the region-of-interest by discriminating between sensor elements that output noise and flow-induced signals. The thermopile array yielded an average root mean squared error of . Parameters such as distance, breathing rate, orientation, and oral or nasal breathing resulted in being fundamental in the detection of respiratory flow. The paper provides the proof-of-concept that low-cost thermopile-arrays can be used to monitor respiratory flow in a lab setting and without the need for facial landmark detection. Further development could provide a more attractive alternative for the earlier bolometer-based proposals.