Fiber Bragg grating (FBG) sensors fabricated in silica optical fiber (Silica-FBG) have been used to measure the strain of human arteries as pulse wave signals. A variety of vital signs including blood pressure can be derived from these signals. However, silica optical fiber presents a safety risk because it is easily fractured. In this research, an FBG sensor fabricated in plastic optical fiber (POF-FBG) was employed to resolve this problem. Pulse wave signals were measured by POF-FBG and silica-FBG sensors for four subjects. After signal processing, a calibration curve was constructed by partial least squares regression, then blood pressure was calculated from the calibration curve. As a result, the POF-FBG sensor could measure the pulse wave signals with an signal to noise (SN) ratio at least eight times higher than the silica-FBG sensor. Further, the measured signals were substantially similar to those of an acceleration plethysmograph (APG). Blood pressure is measured with low error, but the POF-FBG APG correlation is distributed from 0.54 to 0.72, which is not as high as desired. Based on these results, pulse wave signals should be measured under a wide range of reference blood pressures to confirm the reliability of blood pressure measurement uses POF-FBG sensors.
A small-size, high-precision fiber Bragg grating interrogator was developed for continuous plethysmograph monitoring. The interrogator employs optical edge filters, which were integrated with a broad-band light source and photodetector to demodulate the Bragg wavelength shift. An amplifier circuit was designed to effectively amplify the plethysmograph signal, obtained as a small vibration of optical power on the large offset. The standard deviation of the measured Bragg wavelength was about 0.1 pm. The developed edge filter module and amplifier circuit were encased with a single-board computer and communicated with a laptop computer via Wi-Fi. As a result, the plethysmograph was clearly obtained remotely, indicating the possibility of continuous vital sign measurement.