1 July 2021 in Circulation: Arrhythmia and Electrophysiology
Verified
Journal Article
Closed access
PDF file
HNHiroshi Nakagawa
Hiroshi Nakagawa
AIAtsushi Ikeda
Atsushi Ikeda
TSTushar Sharma
Tushar Sharma
AGAssaf Govari
Assaf Govari
JAJohn Ashton
John Ashton
WJWarren M. Jackman
Warren M. Jackman
Background: With short radiofrequency (RF) applications, tissue temperature continues to rise after RF termination (thermal latency), which may result in lesion growth after RF termination. The purpose was to compare in vivo tissue temperature profile (thermal latency), lesion size, and the incidence of steam pop and thrombus between RF ablation with very high power–very short RF (90 W/4 s), high power–short RF (50 W/10 s), and moderate power–moderate RF (30 W/30 s) in a canine thigh muscle preparation and beating heart. Methods: In the thigh muscle preparation (5 dogs), a 3.5-mm ablation electrode with 66 or 56 small irrigation holes (QDOT-Micro or ThermoCoolSmartTouch-SF, respectively) was held perpendicular or parallel to the muscle at 10 g or 30 g contact force. Total of 120 RFs were delivered at 90 W/4 s (QDOT catheter), 50 W/10 s, or 30 W/30 s (SF catheter). Electrode temperature, electrode-tissue interface temperature, and tissue temperatures at 3- and 7-mm depths were measured. In 6 closed-chest dogs, total of 72 RFs were delivered in the ventricle at 90 W/4 s, 50 W/10 s, or 30 W/30 s. Results: In the thigh muscle preparation, tissue temperatures and lesion size (depth, diameter, and volume) were the lowest/smallest for RFs at 90 W/4 s, followed by 50 W/10 s, and the greatest for 30 W/30 s. Thermal latency (Δtemperature and duration) was the greatest for RFs at 90 W/4 s, followed by 50 W/10 s, and the smallest for 30 W/30 s ( P <0.01). Effective tissue heating (area under the curve ≥50 °C at 3-mm depth) was observed after RF termination in 88.0±7.6% with 90 W/4 s, 57.7±14.6% with 50 W/10 s, and only 31.9±8.5% with 30 W/30 s ( P <0.01). In beating hearts, lesion size was also the smallest with 90 W/4 s and the greatest with 30 W/30 s RFs. Increasing contact force significantly increased lesion depth in all 3 groups. There was no significant difference in the incidence of steam pop or thrombus between 3 groups. Conclusions: Tissue temperatures and lesion size (depth, diameter, and volume) were the lowest/smallest for RF applications at 90 W/4 s, followed by 50 W/10 s, and the greatest for 30 W/30 s. The greater thermal latency for 90 W/4 s RF applications suggests that a significant portion of lesion is created after RF termination due to conductive tissue heating.

1 August 2024 in Circulation: Arrhythmia and Electrophysiology
Verified
Journal Article
Closed access
HNHiroshi Nakagawa
Hiroshi Nakagawa
SFSalman Farshchi-Heydari
Salman Farshchi-Heydari
JMJennifer Maffre
Jennifer Maffre
TSTushar Sharma
Tushar Sharma
AGAssaf Govari
Assaf Govari
OWOussama M. Wazni
Oussama M. Wazni
BACKGROUND: During pulsed field ablation (PFA), relationships between ablation parameters (contact force [CF], number of burst pulses, impedance decrease, and electrode temperature) and lesion size in beating hearts have not been well validated. METHODS: A 7.5F-catheter with a 3.5-mm ablation electrode and CF sensor (ThermoCool SmartTouch SF-Dual-Energy, Biosense Webster, Inc, Irwindale, CA) was connected to a PFA system (TRUPULSE2, Biosense Webster, Inc). In 11 closed-chest swine, biphasic PFA current was delivered between the ablation electrode and the skin patch at 219 sites in left ventricle and right ventricle using 12, 18, and 24 burst pulses with 4 different levels of CF: (1) low (n=57; CF, 4–15 g ; median, 10 g ); (2) moderate (n=60; CF, 16–30 g ; median, 22.5 g ); (3) high (n=68; CF, 32–65 g ; median, 40 g ); and (4) no electrode contact (n=34), 2 mm away from the endocardium. Swine were euthanized 2 hours after ablation, and lesion size was measured using triphenyl tetrazolium chloride staining. RESULTS: All PFA lesions with electrode-myocardium contact were well demarcated with triphenyl tetrazolium chloride staining, demonstrating (1) pale central zone (contraction band necrosis with minimal coagulation necrosis), (2) dark brown zone (contraction band necrosis with hemorrhage), and (3) hyperstained red zone by triphenyl tetrazolium chloride (unaffected normal myocardium with preserved mitochondrial activity, consistent with reversible zone). Lesion depth increased significantly with increasing CF and the number of PFA burst pulses. An exponential/logarithmic formula combined with CF and the number of PFA burst pulses correlated lesion depth with high accuracy: R=0.809, P <0.0001, ±1.0-mm accuracy in 128 of 163 (79%) lesions, and ±1.5-mm accuracy in 153 of 163 (94%) lesions. Impedance decrease and electrode temperature were poor predictors of lesion size. There were no detectable lesions resulting from ablation without electrode contact. CONCLUSIONS: Acute PFA ventricular lesions demonstrate irreversible and reversible lesion boundaries. Electrode-tissue contact is required for effective lesion formation. Lesion depth increases significantly with increasing CF and PFA burst pulses. A new exponential/logarithmic formula combined with CF and the number of PFA burst pulses correlates lesion depth with high accuracy.

1 February 2025 in Circulation: Arrhythmia and Electrophysiology
Verified
Journal Article
Open access
AVAtul Verma
Atul Verma
JMJennifer Maffre
Jennifer Maffre
TSTushar Sharma
Tushar Sharma
SFSalman Farshchi-Heydari
Salman Farshchi-Heydari
BACKGROUND: Sequential application of radiofrequency with pulsed field (PF) ablation may increase lesion depth while preserving the advantages of PF. The study’s aim was to determine lesion dimensions of sequential, colocalized radiofrequency and PF ablation. METHODS: A preclinical study using swine (n=4) performed lesions in the right/left ventricles. Ablations were performed with a force-sensing 3.5-mm irrigated-tip ablation catheter using a generator delivering both radiofrequency and PF. PF was delivered using unipolar, biphasic pulses at a standard dose (PF index, 300) with 4-mL/min irrigation. Radiofrequency was delivered at 50 W for 10 s (15 mL/min). Lesions were created by applying colocalized radiofrequency followed by sequential application of PF on the same location, PF followed by sequential application of radiofrequency on the same location, PF alone, or radiofrequency alone. Tissue was collected after 2 hours for lesion assessment. Results are mean±SD. RESULTS: Forty-five lesions were analyzed. The lesion depth of radiofrequency alone was 4.9±0.8 mm. The mean lesion depth and width for PF alone were 3.5±0.6 and 5.1±1.8 mm. Lesion depths for combined applications were significantly greater versus PF alone (6.2±1.8 mm radiofrequency followed by sequential application of PF on the same location; 5.7±1.3 mm PF followed by sequential application of radiofrequency on the same location; P <0.0001 for both). Lesion widths were also significantly greater with combined therapy versus PF alone (8.6±1.8 mm radiofrequency followed by sequential application of PF on the same location; 8.9±2.1 mm PF followed by sequential application of radiofrequency on the same location; P <0.0001 for both). Histology for both combined lesions showed central thermal necrosis surrounded by a hemorrhagic and transitional PF zone. CONCLUSIONS: Combined, colocalized radiofrequency and PF, irrespective of order, show significantly increased lesion size compared with the same dose of PF or radiofrequency alone.

1 December 2023 in Circulation: Arrhythmia and Electrophysiology
Verified
Journal Article
Open access
LBLuigi Di Biase
Luigi Di Biase
JMJacopo Marazzato
Jacopo Marazzato
FZFengwei Zou
Fengwei Zou
ALAung Lin
Aung Lin
VGVito Grupposo
Vito Grupposo
XZXiaodong Zhang
Xiaodong Zhang
Background: Pulsed field ablation (PFA) has emerged as an alternative to radiofrequency ablation. However, data on focal point-by-point PFA are scarce. The aim of this study was to compare lesion durability and collateral damage between focally delivered unipolar/biphasic PFA versus radiofrequency in swine. Methods: Eighteen swine were randomized to low-dose PFA, high-dose PFA, and radiofrequency using a multimodality generator. Radiofrequency delivered by market-available generator served as control group. A contact force–sensing catheter was used to focally deliver PFA/radiofrequency at the pulmonary veins and other predefined sites in the atria. Animals were remapped postprocedurally and 28 days postablation to test lesion durability followed by gross necroscopy and histology. Results: All targeted sites were successfully ablated (contact force value, 13.9±4.1 g). Follow-up remapping showed persistent pulmonary vein isolation in all animals (100%) with lesion durability at nonpulmonary vein sites proven in most (98%). Regardless of the energy source used, the lesion size was similar across the study groups. Transmurality was achieved in 95% of targeted sites and 100% at pulmonary veins. On histology, PFA animals showed more mature scar formation than their radiofrequency counterpart without myocardial necrosis or inflammation. Finally, no sign of collateral damage was observed in any of the groups. Conclusions: In a randomized preclinical study, focally delivered unipolar/biphasic PFA guided by contact force values was associated with durable lesions on chronic remapping and with mature scar formation on histology without signs of collateral injury on necroscopy. Further studies are needed to investigate the long-term feasibility of this new approach to atrial fibrillation treatment.

1 January 2025 in Circulation: Arrhythmia and Electrophysiology
Verified
Journal Article
Closed access
IKIwanari Kawamura
Iwanari Kawamura
RBRahul Bhardwaj
Rahul Bhardwaj
AGAssaf Govari
Assaf Govari
CBChristopher T. Beeckler
Christopher T. Beeckler
JMJennifer Maffre
Jennifer Maffre
JKJacob S. Koruth
Jacob S. Koruth

1 December 2025 in Circulation: Arrhythmia and Electrophysiology
Verified
Journal Article
Open access
HNHiroshi Nakagawa
Hiroshi Nakagawa
SFSalman Farshchi-Heydari
Salman Farshchi-Heydari
MSMasafumi Sugawara
Masafumi Sugawara
AIAtsushi Ikeda
Atsushi Ikeda
JMJennifer Maffre
Jennifer Maffre
OWOussama M. Wazni
Oussama M. Wazni
BACKGROUND: In a previous study, on pulsed-field ablation (PFA) in the swine ventricle, we found that lesion depth was described (±1 mm accuracy) by a logarithmic function of contact force (CF) and the number of PFA pulses (PF-ablation index). This study was designed to validate prospectively whether the novel PF-ablation index would allow PFA lesions to be created at depths of 3.5, 4.5, 5.5, and 6.5 mm with high prediction accuracy in a swine-beating heart model. METHODS: A 7.5F catheter with a 3.5 mm ablation electrode and CF sensor (ThermoCool SmartTouch SF-Dual Energy) was connected to a PFA system (TRUPULSE 2). In 6 closed-chest swine, a biphasic PFA pulse was delivered between the ablation electrode and a skin patch at 123 separate ventricular sites at 5 different levels of CF (1) low (average CF: 4–15 g; median, 12 g; n=25), (2) moderate (16–30 g; median, 23 g; n=41); 3) high (31–45 g; median, 36 g; n=27), (4) very high (46–68 g; median, 52 g; n=18); or (5) no electrode contact, 1 to 2 mm from the endocardium (n=12). PFA application was terminated when the PF-ablation index reached a predicted lesion depth of 3.5 mm (27 sites), 4.5 mm (25 sites), 5.5 mm (29 sites), and 6.5 mm (30 sites). Swine were euthanized 2 hours after ablation. Lesion size was measured using triphenyl tetrazolium chloride staining. RESULTS: Predicted lesion depth by the PF-ablation index correlated well with actual lesion depthwith ±1.0 mm accuracy in 97/106 (92%) lesions and ±1.5 mm accuracy in all 106 lesions. There were no or poor relationships between intracardiac electrogram attenuation, impedance decrease, electrode temperature, and lesion size. No detectable lesions were created without electrode contact. CONCLUSIONS: A novel PF-ablation index incorporating CF and the number of PFA pulses provides high accuracy in predicting lesion depth in real-time. Intracardiac electrogram attenuation, impedance decrease, and electrode temperature are poor predictors of PFA lesion size.