The stimulation of deeper brain regions with transcranial magnetic stimulation (TMS) is a promising treatment for depression with higher remission rates. The focality of the stimulation area is essential to reduce the risk of side effects; however, there is a physical tradeoff between depth and focality. Because existent studies have searched for depth by iteratively designing a coil and then calculating the electric field in the brain, the theoretical maximum depth has not yet been determined. In this study, we propose a methodology to derive the coil based on an inverse problem approach from an analytical description of the desired electric field in the brain. This new method has allowed us to theoretically derive the maximum depth of the electric field, which was previously only determined empirically. Based on this, we designed and built a coil that can deliver the electric field to the deepest depth. Numerical simulations of the proposed coils show that they have the highest electric field depth of any coil feasible in the existing medical field. Moreover, the balance between depth and focality was also the best of the existing coils.
Trans-spinal magnetic stimulation (TSMS) is a non-invasive method to stimulate nerves in the spinal cord and has recently garnered attention. However, delivering sufficient stimulation while suppressing pain during stimulation remains challenging because nerves in the spinal cord are located deep within the body and surrounded by other tissues including high-impedance structures. This study designed an optimal TSMS coil incorporating the electrical conductivities of biological tissues via numerical simulations based on a digital human model. The stream function representing the coil geometry was expanded into a Fourier series and each basis function was transformed into an induced electric field via simulations. The induced electric field was subsequently expressed as a linear combination of the basis electric fields. An optimal induced electric field was derived that is sufficient to stimulate spinal cord nerves while being suppressed at the body surface. The optimal coil geometry was obtained by optimizing the basis coefficients. The induced electric field generated by the proposed coil achieved an improved penetration depth by 19% and higher field strength by about 3 V/m compared to that of the conventional figure-8 coil. The proposed coil is expected to provide a modest improvement within the induced electric field in the spinal cord region while reducing the field at the body surface.