A body pressure relief system was newly developed with optical pressure sensors for pressure ulcer prevention. Unlike a conventional alternating pressure air mattress (APAM), this system automatically regulates air flow into a body supporting mattress with adaptive inflation (or deflation) duration in response to the pressure level in order to reduce skin stress due to prolonged high pressures. The system continuously quantifies the body pressure distribution using time-of-flight (ToF) optical sensors. The proposed pressure sensor, a ToF optical sensor in the air-filled cell, measures changes in surface height of mattress when pressed under body weight, thereby indirectly indicating the interface pressure. Non-contact measurement of optical sensor usually improves the durability and repeatability of the system. The pressure sensor was successfully identified the 4 different-predefined postures, and quantitatively measured the body pressure distribution of them. Duty cycle of switches in solenoid valves was adjusted to 0–50% for pressure relief, which shows that the interface pressure was lower than 32 mmHg for pressure ulcer prevention.
Amputees typically experience changes in residual limb volume in their daily lives. It causes an uncomfortable fit of the socket by applying high pressure on the sensitive area of the residual limb or by loosening the socket. In this study, we developed a transfemoral prosthetic socket for above-the-knee amputees that ensures a good socket fit by maintaining uniform and constant contact pressure despite volume changes in the residual limb. The socket has two air bladders in the posterior femoral region, and the pneumatic controller is located on the tibia of the prosthesis. The pneumatic system aims to minimize unstable fitting of the socket and improve walking performance by inflating or deflating the air bladder. The developed socket autonomously maintains the air pressure inside the prosthetic socket at a steady-state error of 3 mmHg or less by adjusting the amount of air in the air bladder via closed-loop control. In the clinical trial, amputee participants walked on flat and inclined surfaces. The displacement between the residual limb and socket during the gait cycle was reduced by up to 33.4% after air injection into the socket. The inflatable bladder increased the knee flexion angle on the affected side, resulting in increased stride length and gait velocity. The pneumatic socket provides a stable and comfortable walking experience not only when walking on flat ground but also on slopes.