The biplane coils (BCs) based on the inverse-design method are crucial for creating the near-zero magnetic environment in magnetic shielding room (MSR), but multi-parameters optimization remains a key factor that limits the performance of the BCs. This study proposes a funnel algorithm (FA) to solve the issue of multi-parameter optimization in the design. The upper and lower limits of the parameters are dynamically adjusted based on the previous calculating step. By gradually reducing the upper and lower limits, we achieve the BCs with optimal performance. This method has been verified by both the theoretical analysis and experimental test. Compared with other methods, the designed coils show a higher space utilization ratio (SUR) and smaller relative errors, with the SUR improved from 1.6% to 3.75% and the relative errors all less than 2%. Based on the designed BCs, the total residual magnetic field is stabilized below 40.4 pT at the central point of the MSR, with the noise reaching the intrinsic noise of the flux-gate magnetometer in the frequency range of interest. The proposed method can significantly improve the performance of the BCs, thus reducing the space and construct cost of the MSR and promoting the magnetoencephalogram (MEG) measurement based on the quantum technology.
An active magnetic compensation method based on passive shielding is used for low remanence and interfering field suppression in a lower cost magnetic shielding room (MSR). In this paper, a compensation coil control strategy with conventional PI and fuzzy PI segmentation control is proposed, and an active magnetic compensation system with a linearized model is built using an external high uniformity Helmholtz coil. The result shows that the compensation system has a greatly improvement in the suppression of the interfering magnetic field in the environment while ensuring good step response performance of the controller.
28 March 2025 in IEEE Transactions on Instrumentation and Measurement
A large uniform region of magnetic field-free space is of great importance for magnetoencephalography (MEG) in a dynamic environment. Existing active compensation systems only compensate the magnetic field in a relatively small uniform region, the space utilization ratio (SUR) is low, as limited by the design theory. In this paper, the design and the test of spatial constraint coil (SCC) are introduced. This type of coil maintains high uniformity in a relatively large region after special optimization, the performance test results show that the SUR of the SCC can reach 12.5 %, which is an improvement of about 10 times compared with existing compensation coils. A magnetic field compensation system based on the SCC has been developed. After compensation, the residual magnetic field in our magnetic shielded room (MSR) is reduced to 0.5 nT in all three axes, which is 0.4 % of the residual field before compensation, and the noise level is below 10 pT / Hz1/2 at 0.1-10 Hz. This method can greatly increase the uniform region while keeping high uniformity and compensation accuracy, which will provide a new solution for the creation of a large uniform region of magnetic field-free space in dynamic MEG measurement.
25 March 2026 in IEEE Transactions on Industrial Electronics
In complex magnetic environments, accurately estimating and suppressing multiple uncertain periodic and aperiodic disturbances is critical for magnetocardiography (MCG) imaging. This article proposes an adaptive active disturbance rejection controller (ADRC) with a multiresonator and a variable-step multifrequency estimator for the active magnetic field control (AMFC) system. The core innovation is the variable-step multifrequency estimator that enables real-time, precise, and adaptive tracking of multiple disturbance frequencies. Integrated with the multiresonator extended state observer (ESO), the tracked frequencies are dynamically allocated to the corresponding resonant center frequencies, allowing simultaneous observation and suppression of both multiple periodic and aperiodic magnetic disturbances. Experimental validation within a magnetically shielded room (MSR) demonstrates that the proposed method significantly outperforms conventional fixed-step methods, achieving: 1) a 21.2% faster convergence time and a 56.3% lower frequency estimation error compared to fixed-step estimators; 2) a 73.3% higher disturbance suppression ratio; and 3) a 10.2% improvement in the signal-to-noise ratio (SNR) of MCG signals. These results highlight the superior performance of the proposed method in precise magnetic field control for MCG.