The design, fabrication, assembly and testing of a modular dynamic torso phantom for magnetic resonance imaging (MRI) is presented. The phantom is composed of fully MRI-compatible materials and consists of a saline-filled acrylic shell, with dimensions comparable to a human torso, and three distinct interior motion modules. Two linear piezo stages are used to mimic respiratory motion of the chest-wall (≤ 10 mm; cycle time ≥ 2 s) and the abdominal organs (≤ 45 mm; cycle time ≥ 2 s). A pneumatic system is used to move a heart mock-up to represent cardiac motion (≤ 19 mm, rate ≤ 3 Hz). The motion capabilities and MRI compatibility of the phantom are tested in 3 T MRI experiments. The closed-loop control of the phantom’s breathing modules allows for accurate positioning and timing with absolute differences between input and output values ≤ 0.61 mm and ≤ 0.06 s. The purpose of this phantom is to serve as a scientific instrument for the future development and evaluation of motion tracking and artifact compensation techniques.