Radiation detectors have been fabricated from Cd1-xZnxTe crystals and their performance has been studied with respect their Cd1-xZnxTe surface preparation. Energy resolution, leakage current, resistivity, ohmicity as well as surface roughness of the detectors were measured and compared for various chemical treatments. Etching in lactic acid has been found to lead to better ohmic contract, a rougher surface and higher leakage current. Bromine is effective in removing the damaged layer and when added to 20 percent lactic acid in ethylene glycol leads to a lower leakage current and consequently a better energy resolution. The best detector performance was obtained with a two step treatment consisting of a standard etchant (5 percent Br in methanol) followed by 2 percent Br in'20 percent lactic acid in ethylene glycol'. The results clearly demonstrate the important of surface chemical treatments in the fabrication of Cd1-xZnxTe radiation detectors.
The High Efficiency Multimode Imager (HEMI) is an instrument to detect, locate, and spectroscopically characterize radioactive sources with gamma-ray emissions at long-range standoff distances. HEMI consists of modular cubic-centimeter coplanar-grid CdZnTe detector elements configured in a two-plane array that allows for both Compton scattering and coded aperture imaging. Measurements and simulations have been performed using a variety of radioactive sources at different positions with respect to a prototype HEMI array in order to demonstrate the source identification and localization capabilities of the system. This paper will provide an overview of the HEMI instrument as well as present results from simulations and measurements using the completed detector array.