An imaging platform based on broadband coherent anti-Stokes Raman scattering has been developed that provides an advantageous combination of speed, sensitivity and spectral breadth. The system utilizes a configuration of laser sources that probes the entire biologically relevant Raman window (500–3,500 cm–1) with high resolution (<10 cm–1). It strongly and efficiently stimulates Raman transitions within the typically weak ‘fingerprint’ region using intrapulse three-colour excitation, and utilizes the non-resonant background to heterodyne-amplify weak Raman signals. We demonstrate high-speed chemical imaging in two- and three-dimensional views of healthy murine liver and pancreas tissues as well as interfaces between xenograft brain tumours and the surrounding healthy brain matter.
We describe a closed-form approach for performing a Kramers–Kronig (KK) transform that can be used to rapidly and reliably retrieve the phase, and thus the resonant imaginary component, from a broadband coherent anti-Stokes Raman scattering (CARS) spectrum with a nonflat background. In this approach we transform the frequency-domain data to the time domain, perform an operation that ensures a causality criterion is met, then transform back to the frequency domain. The fact that this method handles causality in the time domain allows us to conveniently account for spectrally varying nonresonant background from CARS as a response function with a finite rise time. A phase error accompanies KK transform of data with finite frequency range. In examples shown here, that phase error leads to small errors in the retrieved resonant spectra.