Michael R. Bruesewitz

Michael R. Bruesewitz

Auto-generated profile, based on publicly available information

Michael R. Bruesewitz

From the Department of Radiology (Shaojie Chang, John C. Benson, John I. Lane, Michael R Bruesewitz, Joseph R Swicklik, Jamison E. Thorne, Emily K. Koons, Cynthia H. McCollough, Shuai Leng) and Department of Otolaryngology–Head and Neck Surgery (Matthew L. Carlson), Mayo Clinic, Rochester, MN, 55905, USA

Research Focus & Metrics

Academic Career Length

9Years (2018-Present)

(Calculated once Scilit publication data is synced: current year − year of first publication + 1)

Representative Publications

View All Publications
9 August 2018 in American Journal of Neuroradiology
Verified
Journal Article
Open access
PDF file
WZW. Zhou
W. Zhou
JLJohn Lane
John Lane
Matthew CarlsonMatthew Carlson
Matthew Carlson
MBM.R. Bruesewitz
M.R. Bruesewitz
RWR.J. Witte
R.J. Witte
SLS. Leng
S. Leng
BACKGROUND AND PURPOSE: Evaluating abnormalities of the temporal bone requires high-spatial-resolution CT imaging. Our aim was to assess the performance of photon-counting-detector ultra-high-resolution acquisitions for temporal bone imaging and compare the results with those of energy-integrating-detector ultra-high-resolution acquisitions. MATERIALS AND METHODS: Phantom studies were conducted to quantify spatial resolution of the ultra-high-resolution mode on a prototype photon-counting-detector CT scanner and an energy-integrating-detector CT scanner that uses a comb filter. Ten cadaveric temporal bones were scanned on both systems with the radiation dose matched to that of the clinical examinations. Images were reconstructed using a sharp kernel, 0.6-mm (minimum) thickness for energy-integrating-detector CT, and 0.6- and 0.25-mm (minimum) thicknesses for photon-counting-detector CT. Image noise was measured and compared using adjusted 1-way ANOVA. Images were reviewed blindly by 3 neuroradiologists to assess the incudomallear joint, stapes footplate, modiolus, and overall image quality. The ranking results for each specimen and protocol were compared using the Friedman test. The Krippendorff α was used for interreader agreement. RESULTS: Photon-counting-detector CT showed an increase of in-plane resolution compared with energy-integrating-detector CT. At the same thickness (0.6 mm), images from photon-counting-detector CT had significantly lower (P < .001) image noise compared with energy-integrating-detector CT. Readers preferred the photon-counting-detector CT images to the energy-integrating-detector images for all 3 temporal bone structures. A moderate interreader agreement was observed with the Krippendorff α = 0.50. For overall image quality, photon-counting-detector CT image sets were ranked significantly higher than images from energy-integrating-detector CT (P < .001). CONCLUSIONS: This study demonstrated substantially better delineation of fine anatomy for the temporal bones scanned with the ultra-high-resolution mode of photon-counting-detector CT compared with the ultra-high-resolution mode of a commercial energy-integrating-detector CT scanner.

17 December 2020 in Medical Physics
Verified
Journal Article
Open access
PDF file
STShengzhen Tao
Shengzhen Tao
ESEmily Sheedy
Emily Sheedy
MBMichael Bruesewitz
Michael Bruesewitz
NWNikkole Weber
Nikkole Weber
KWKyle Williams
Kyle Williams
SLShuai Leng
Shuai Leng
Purpose: To determine the accuracy of CT number and calcium score of a kV‐independent technique based on an artificial 120 kV reconstruction, and its potential to reduce radiation dose.Methods: Anthropomorphic chest phantoms were scanned on a third‐generation dual‐source CT system equipped with the artificial 120 kV reconstruction. First, a phantom module containing a 20‐mm diameter hydroxyapatite (HA) insert was scanned inside the chest phantoms at different tube potentials (70–140 kV) to evaluate calcium CT number accuracy. Next, three small HA inserts (diameter/length = 5 mm) were inserted into a pork steak and scanned inside the phantoms to evaluate calcium score accuracy at different kVs. Finally, the same setup was scanned using automatic exposure control (AEC) at 120 kV, and then with automatic kV selection (auto‐kV). Phantoms were also scanned at 120 kV using a size‐dependent mA chart. CT numbers of soft tissue and calcium were measured from different kV images. Calcium score of each small HA insert was measured using commercial software.Results: The CT number difference from 120 kV was small with tube potentials from 90 to 140 kV for both soft tissue and calcium (maximal difference of 4/5 HU, respectively). Consistent calcium scores were obtained from images of different kVs compared to 120 kV, with a relative difference <8%. Auto‐kV provided a 25–34% dose reduction compared to AEC alone.Conclusion: A kV‐independent calcium scoring technique can produce artificial 120 kV images with consistent soft tissue and calcium CT numbers compared to standard 120 kV examinations. When coupled with auto‐kV, this technique can reduce radiation by 25–34% compared to that with AEC alone, while providing consistent calcium scores as that of standard 120 kV examinations.

11 November 2024 in American Journal of Neuroradiology
Verified
Journal Article
Open access
PDF file
SCShaojie Chang
Shaojie Chang
JBJohn C. Benson
John C. Benson
JLJohn Lane
John Lane
MBMichael R. Bruesewitz
Michael R. Bruesewitz
JSJoseph R. Swicklik
Joseph R. Swicklik
SLShuai Leng
Shuai Leng
BACKGROUND AND PURPOSE: Ultra-high-resolution (UHR) photon-counting-detector (PCD) CT improves image resolution but increases noise, necessitating use of smoother reconstruction kernels that reduce resolution below the system’s 0.110 mm maximum spatial resolution. To address this, a denoising convolutional neural network (CNN) was developed to reduce noise in images reconstructed with the available sharpest reconstruction kernel while preserving resolution for enhanced temporal bone visualization. MATERIALS AND METHODS: With IRB approval, CNN was trained on 6 clinical temporal bone patient cases (1,885 images) and tested on 20 independent cases using a dual-source PCD-CT (NAEOTOM Alpha, Siemens). Images were reconstructed using iterative reconstruction at strength 3 (QIR3) with both clinical routine (Hr84) and the sharpest available head kernel (Hr96). The CNN was applied to images reconstructed with Hr96 and QIR1. Three image series (Hr84-QIR3, Hr96-QIR3, and Hr96-CNN) for each case were randomized for review by two neuroradiologists, assessing overall quality and delineation of the modiolus, stapes footplate, and incudomallear joint. RESULTS: CNN reduced noise by 80% compared to Hr96-QIR3 and 50% relative to Hr84-QIR3, while maintaining high resolution. When compared to the conventional method at the same kernel (Hr96-QIR3), Hr96-CNN significantly decreased image noise (from 204.63 HU to 47.35 HU) and improved SSIM (from 0.72 to 0.99). Hr96-CNN images ranked higher than Hr84-QIR3 and Hr96-QIR3 in overall quality (p<0.001). Readers preferred Hr96-CNN for all three structures. CONCLUSIONS: The proposed CNN significantly reduced image noise in UHR PCD-CT, enabling the use of sharpest kernel. This combination greatly enhanced diagnostic image quality and anatomical visualization. ABBREVIATIONS: PCD = Photon-counting-detector; UHR = Ultra-high-resolution; IR = Iterative reconstruction; CNN = Convolutional neural network; SSIM: Structural similarity index.