4 December 2025 in IEEE Sensors Journal
An adaptive algorithm is proposed for selecting the decomposition level of intrinsic mode functions (IMFs) that contain vital signals in a non-contact bio-radar sensing system. Mode decomposition techniques are widely employed to filter out high-frequency noise from radar signals, enabling effective extraction of respiratory and cardiac components. However, the IMF levels containing critical biological information vary due to environmental noise and individual subject differences. The proposed algorithm accurately identifies the appropriate IMF levels for respiratory and cardiac signals without requiring prior knowledge of the input signal. Specifically, it effectively distinguishes vital signs from unwanted noise components, such as respiratory harmonics, that may overlap in frequency bands. This is achieved by estimating significant frequency ranges within each IMF based on analysis of their phase spectrum characteristics. Additionally, the algorithm analyzes offset patterns in the phase components to determine precise frequency boundaries and selects relevant IMF levels accordingly. Experimental evaluations conducted with continuous measurement data from an in-house 5.8-GHz bio-radar sensor demonstrate that the proposed method flexibly adapts to changing IMF levels corresponding to respiratory and cardiac signal. Compared to conventional signal-to-noise ratio-based IMF selection methods, the proposed algorithm reduces the average root mean square error for respiratory rate and heart rate measurements by 9.3 bpm and 3.5 bpm, respectively. This corresponds to improvements in measurement accuracy of 3.3% for respiration and 11.6% for heart rate.