<!-- *** Custom HTML *** -->Optical-resolution photoacoustic microscopy (OR-PAM) is applied to functional brain imaging in living mice. A near-diffraction-limited bright-field optical illumination is employed to achieve micrometer lateral resolution, and a dual-wavelength measurement is utilized to extract the blood oxygenation information. The variation in hemoglobin oxygen saturation (sO
2) along vascular branching has been imaged in a precapillary arteriolar tree and a postcapillary venular tree, respectively. To the best of our knowledge, this is the first report on in vivo volumetric imaging of brain microvascular morphology and oxygenation down to single capillaries through intact mouse skulls. It is anticipated that: (
i) chronic imaging enabled by this minimally invasive procedure will advance the study of cortical plasticity and neurological diseases; (
ii) revealing the neuroactivity-dependent changes in hemoglobin concentration and oxygenation will facilitate the understanding of neurovascular coupling at the capillary level; and (
iii) combining functional OR-PAM and high-resolution blood flowmetry will have the potential to explore cellular pathways of brain energy metabolism. Advances in brain imaging facilitate the understanding of cognitive phenomena and neurological diseases. However, high-resolution brain imaging through intact animal skulls remains challenging for pure optical modalities because the optical scattering and absorption of the skull degrade the imaging resolution and signal-to-noise ratio (SNR). Photoacoustic imaging, combining light and ultrasound in a single hybrid technology, suggests a potential solution. Using acoustic-resolution photoacoustic microscopy (AR-PAM), Stein recently demonstrated mouse brain imaging through both intact scalp and skull.
1 This noninvasive feature is highly desirable for functional or chronic studies; however, with its current spatial resolution (lateral resolution:
; axial resolution:
), capillaries are not resolvable. To fill the gap, Maslov developed optical-resolution photoacoustic microscopy (OR-PAM) capable of imaging single capillaries
in vivo.
2 The lateral resolution of OR-PAM matches the size of a single red blood cell (RBC), and its sensitivity enables single RBC detection.
3 Here, we report on the first demonstration of OR-PAM for functional brain microvascular imaging down to single capillaries through intact mouse skulls. The minimally invasive feature is favorable for chronic study of cortical plasticity. Moreover, because neuronal activity is widely assumed to spatially correlate most closely to the capillary bed response,
4 improving localization of signals down to the capillary level will enable functional brain mapping at micrometer resolution. Before functional brain imaging, a Swiss Webster mouse (Hsd:ND4, Harlan Co.,
) was anesthetized by intraperitoneally administering a dose of
ketamine and
xylazine and transferred to a stereotaxic imaging stage. The scalp of the mouse was surgically removed, and the exposed skull was cleaned with 0.9% sodium chloride irrigation solution right before imaging. Ultrasonic gel was used for ultrasound coupling and maintaining skull hydration. Throughout the experiment, the animal was supplied with breathing-grade air and maintained under anesthesia using vaporized isoflurane (1.0–1.5% isoflurane with an airflow rate of
). The body temperature of the animal was maintained at
by a temperature-controlled heating pad. At the end of the experiment protocol, the animal was euthanatized by an intraperitoneal administration of pentobarbital at a dosage of
. The detailed system description can be found in our published papers.
2, 3 In photoacoustic measurements of hemoglobin concentration and oxygenation, we assume that, in the visible spectral range, oxyhemoglobin
and deoxyhemoglobin (HbR) are the dominant absorbing compounds in blood.
5 Thus, a dual-wavelength measurement is adequate to image
, though using more wavelengths is expected to yield more accurate results.
6 The two wavelengths chosen here are 570 and
, where the absorption contrast between blood and background brain tissues are high enough to enable satisfactory imaging quality.
5 According to the published absorption spectra of rat hemoglobin,
7 is an isosbestic point, at which
and HbR have the same molar extinction coefficients. Thus, the photoacoustic signal acquired at this wavelength reflects the total hemoglobin concentration (HbT), and
is an
-absorption-dominant wavelength, which helps differentiate the two types of hemoglobin. Because blood oxygenation is highly correlated with local metabolism, the
value is generally time variant. To minimize measurement error due to the possible temporal fluctuation in blood oxygenation, we implemented a wavelength autotuning program to control the dye laser, and imaged the same cross-sectional scan (B-scan) for each of the wavelengths before moving to the next B-scan. The region of interest (ROI) was scanned with a step size of
. The dual-wavelength measurement took
, which was mostly limited by the wavelength tuning speed of the dye laser. Figures 1a and 1b are maximum amplitude projection (MAP) images of the mouse brain microvasculature under systemic normoxia at the optical wavelengths of 570 and
, respectively. Because
is an isosbestic point, Fig. 1a maps the HbT, regardless of the blood oxygenation. Microvessels labeled with CL in Fig. 1a appear to be single capillaries with diameters of
. A corresponding volumetric rendering is shown in Videos 1 . At the conclusion of the dual-wavelength measurement, the concentrations of HbR and
, as well as the
values, were calculated based on the model described in previously published work.
5, 6 As shown in Fig. 1c, different
levels are visualized with pseudocolors ranging from blue to red, while the HbT measured at the isosbestic point
is represented by pixel brightness. According to known physiology, the red vessels (
values of
) are believed to be arterioles, whereas the green ones (
values as low as 60–70%) are most likely to be venules. However, in the microcirculation, there is no clear cutoff in
value between arterioles and venules.
8 The blood oxygenation is closely associated with the microvascular branching order
8 and the local metabolic activity of the tissue.
9 To demonstrate the oxygen gradients in the brain microcirculation, we selectively analyzed a postcapillary venular tree [Fig. 1d] and a precapillary arteriolar tree [Fig. 1e] highlighted in Fig. 1c by the blue and red dashed boxes, respectively. The vessel diameters and the corresponding
values in different branching orders are listed in Tables 1, 2 . A negative correlation between the branching order and its mean
value is observed in both the arteriolar and venular trees with a linear regression analysis (arteriolar tree:
,
venular tree:
,
, as shown in Fig. 2 . Our results suggest that: (
i) the blood oxygenation level is higher in the arteriolar tree than in the venular tree; (
ii) capillaries are not the only oxygen exchange site in the microcirculation because the
decreases significantly with vascular branching in the precapillary arteriolar tree (
from order 1 to 4); and (
iii) a diffusional shunt is present between arterioles and venules to elevate the oxygen level in “large” venules because the
increases noticeably with blood confluence in the postcapillary venular tree (
from order 1 to 4). Our observation is in agreement with the published work.
8, 9, 10 <!-- named anchor --> Fig. 1
In vivo functional OR-PAM imaging of mouse brain microvasculature through an intact skull. (a) MAP image acquired at
; (b) MAP image acquired at
; (c) vessel-by-vessel
mapping; (d) a venular tree with branching orders [boxed by blue dashed lines in (c)]; (e) an arteriolar tree with branching orders [boxed by red dashed lines in (c)]. The calculated
values are shown in the color bar. PA: photoacoustic signal amplitude. The scale bar applies to (a), (b), and (c). (Color online only.) <!-- named anchor --> Fig. 2 The
value versus the vascular branching order in a precapillary arteriolar tree and a postcapillary venular tree. The square and round markers represent the mean
values in each branching order, and the error bars stand for the standard deviations of the mean
values. <!-- named anchor --> Table 1 The vessel diameters and the sO2 values in different branching orders in a venular tree. Values are in mean±standard deviation format. Branchingorder 1 2 3 4 Vessel ID V1 V2(a) V2(b) V3(a) V3(b) V3(c) V3(d) V4(a) V4(b) Diameter
(%)
<!-- named anchor --> Table 2 The vessel diameters and the sO2 values in different branching orders in an arteriolar tree. Values are in mean±standard deviation format. Branchingorder 1 2 3 4 Vessel ID A1(a) A1(b) A2(a) A2(b) A3(a) A3(b) A3(c) A3(d) A4(a) A4(b) Diameter
(%)