11 July 2018 in Light: Science & Applications
The common assumption that precision is the limit of accuracy in localization microscopy and the typical absence of comprehensive calibration of optical microscopes lead to a widespread issue—overconfidence in measurement results with nanoscale statistical uncertainties that can be invalid due to microscale systematic errors. In this article, we report a comprehensive solution to this underappreciated problem. We develop arrays of subresolution apertures into the first reference materials that enable localization errors approaching the atomic scale across a submillimeter field. We present novel methods for calibrating our microscope system using aperture arrays and develop aberration corrections that reach the precision limit of our reference materials. We correct and register localization data from multiple colors and test different sources of light emission with equal accuracy, indicating the general applicability of our reference materials and calibration methods. In a first application of our new measurement capability, we introduce the concept of critical-dimension localization microscopy, facilitating tests of nanofabrication processes and quality control of aperture arrays. In a second application, we apply these stable reference materials to answer open questions about the apparent instability of fluorescent nanoparticles that commonly serve as fiducial markers. Our study establishes a foundation for subnanometer localization accuracy in widefield optical microscopy.
24 June 2021 in Nature Communications
A standard paradigm of localization microscopy involves extension from two to three dimensions by engineering information into emitter images, and approximation of errors resulting from the field dependence of optical aberrations. We invert this standard paradigm, introducing the concept of fully exploiting the latent information of intrinsic aberrations by comprehensive calibration of an ordinary microscope, enabling accurate localization of single emitters in three dimensions throughout an ultrawide and deep field. To complete the extraction of spatial information from microscale bodies ranging from imaging substrates to microsystem technologies, we introduce a synergistic concept of the rigid transformation of the positions of multiple emitters in three dimensions, improving precision, testing accuracy, and yielding measurements in six degrees of freedom. Our study illuminates the challenge of aberration effects in localization microscopy, redefines the challenge as an opportunity for accurate, precise, and complete localization, and elucidates the performance and reliability of a complex microelectromechanical system.
18 March 2024 in Optica Quantum
In a popular integration process for quantum information technologies, localization microscopy of quantum emitters guides lithographic placement of photonic structures. However, a complex coupling of microscopy and lithography errors degrades registration accuracy, severely limiting device performance and process yield. We introduce a methodology to solve this widespread but poorly understood problem. A new foundation of traceable localization enables rapid characterization of lithographic standards and comprehensive calibration of cryogenic microscopes, revealing and correcting latent systematic effects. Of particular concern, we discover that scale factor deviation and complex optical distortion couple to dominate registration errors. These novel results parameterize a process model for integrating quantum dots and bullseye resonators, predicting higher yield by orders of magnitude, depending on the Purcell factor threshold as a quantum performance metric. Our foundational methodology is a key enabler of the lab-to-fab transition of quantum information technologies and has broader implications to cryogenic and correlative microscopy.
21 August 2022 in Advanced Functional Materials
Focused‐ion‐beam machining is a powerful process to fabricate complex nanostructures, often through a sacrificial mask that enables milling beyond the resolution limit of the ion beam. However, current understanding of this super‐resolution effect is empirical in the spatial domain and nonexistent in the temporal domain. This article reports the primary study of this fundamental tradespace of resolution and throughput. Chromia functions well as a masking material due to its smooth, uniform, and amorphous structure. An efficient method of in‐line metrology enables characterization of ion‐beam focus by scanning electron microscopy. Fabrication and characterization of complex test structures through chromia and into silica probe the response of the bilayer to a focused beam of gallium cations, demonstrating super‐resolution factors of up to 6 ± 2 and improvements to volume throughput of at least factors of 42 ± 2, with uncertainties denoting 95% coverage intervals. Tractable theory models the essential aspects of the super‐resolution effect for various nanostructures. Application of the new tradespace increases the volume throughput of machining Fresnel lenses by a factor of 75, enabling the introduction of projection standards for optical microscopy. These results enable paradigm shifts of sacrificial masking from empirical to engineering design and from prototyping to manufacturing.
25 October 2018 in Journal of Microelectromechanical Systems
Microelectromechanical systems (MEMS) that require contact of moving parts to implement complex functions exhibit limits to their performance and reliability. Here, we advance our particle tracking method to measure MEMS motion in operando at nanometer, microradian, and millisecond scales. We test a torsional ratcheting actuator and observe dynamic behavior ranging from nearly perfect repeatability, to transient feedback and stiction, to terminal failure. This new measurement capability will help to understand and improve MEMS motion.