18 February 2022 in IEEE Transactions on Instrumentation and Measurement
In a practical multichannel magnetocardiography (MCG) system based on superconducting quantum interference device (SQUID) magnetometers, the Tesla/Volt coefficient must be precisely calibrated for accurate analysis of heart diseases. In this article, we utilize the biplanar coil to calibrate the SQUID magnetometers in an MCG system for the first time. The coil design theory is discussed in detail, including the target-field theory and the Tikhonov regularization method. Also, the coil winding patterns can be obtained after discretization of the stream function, which is related to the current density. By defining the parameter of calibration uncertainty, the performance of the biplanar coil is analyzed and compared with a rectangular Helmholtz coil based on the COMSOL simulation method and experimental measurement. The measurement results are well in coincidence with the simulation ones, indicating that the biplanar coil has remarkable advantages over the Helmholtz coil. Also, Tesla/Volt coefficient calibrations of four SQUID magnetometers in an MCG system are conducted inside a magnetically shielded room (MSR) depending on these two coils. The final results show the superiority of the biplanar coil method in MCG system calibration.
12 January 2024 in IEEE Transactions on Instrumentation and Measurement
In this article, a modified design method is proposed for a kind of highly uniform magnetic field (UMF) coil, whose target region is off-center defined. Combining the traditional target field method (TFM) with the slime mold algorithm (SMA), a new intelligent algorithm with excellent searching ability, optimal parameters can be obtained to determine the structure of the designed coil. To further improve the uniformity performance in the target region, a distance factor is introduced to optimize the distribution of target points. According to the simulation results from the finite element analysis (FEA), the absolute maximum uniformity error can be optimized from 0.098% to 0.060% if the SMA is utilized, and after the optimization of target points, the volume proportion of absolute uniformity error less than $10^{-4}$ can be increased to 64.60%. An upscaled coil of 10 times is fabricated via the flexible printed circuit technology, and test results of the magnetic field generated by this coil show a good agreement with the simulation ones. It indicates that the actual uniformity error in the range of ±0.35 R reaches to 0.11%, which is acceptable when taking manufacture and measurement errors into consideration.
9 April 2024 in IEEE Sensors Journal
Synthetic gradiometers constructed by Superconducting QUantum Interference Device (SQUID) magnetometers are effective ways to suppress the magnetic disturbance while keeping the useful signals unaffected. In this paper, in order to obtain the optimal baseline for a first-order synthetic gradiometer (FOSG) inside a thin magnetically shielding room (MSR), a modeling method based on the magnetic flux distributions in the sensing and reference magnetometers is proposed. Then the model is built in the Ansys platform, and the baseline can be determined as 6 cm under a certain signal source depth of 4 cm after conducting the finite element calculation. Two kinds of magnetometers with different pickup coils contribute to the realization of the testing system. The testing results have a good agreement with the simulation ones. And depending on this optimized configuration of the FOSG, the residual field of a real thin MSR can be restrained from 2.0 nT to 4.0 pT. Moreover, a clear magnetocardiography (MCG) signal with an amplitude of 48 pT and stable zero-line can be observed when applying a subject for test, which further verifies the effectiveness of this modeling method.