To accelerate high-resolution diffusion-weighted imaging with a multi-shot echo-planar sequence, we propose an approach based on reduced averaging and deep learning. Denoising convolutional neural networks can reduce amplified noise without requiring extensive averaging, enabling shorter scan times and high image quality. The preliminary experimental results demonstrate the superior performance of the proposed denoising method over state-of-the-art methods such as the widely used block-matching and 3D filtering.
Purpose: Quantitative volumetricT1mapping in the liver has the potential to aid in the detection, diagnosis, and quantification of liver fibrosis, inflammation, and spatially resolved liver function. However, accurate measurement of hepaticT1is confounded by the presence of fat and inhomogeneous excitation. Furthermore, scan time constraints related to respiratory motion require tradeoffs of reduced volumetric coverage and/or increased acquisition time. This work presents a novel 3D acquisition and estimation method for confounder‐correctedT1measurement over the entire liver within a single breath‐hold through simultaneous estimation ofT1, fat and .Theory and Methods: The proposed method combines chemical shift encoded MRI and variable flip angle MRI with a mapping technique to enable confounder‐correctedT1mapping. The method was evaluated theoretically and demonstrated in both phantom and in vivo acquisitions at 1.5 and 3.0T. At 1.5T, the method was evaluated both pre‐ and post‐ contrast enhancement in healthy volunteers.Results: The proposed method demonstrated excellent linear agreement with reference inversion‐recovery spin‐echo basedT1in phantom acquisitions at both 1.5 and 3.0T, with minimal bias (5.2 and 45 ms, respectively) overT1ranging from 200–1200 ms. In vivo results were in general agreement with reference saturation‐recovery based 2DT1maps (SMART1Map, GE Healthcare).Conclusion: The proposed 3DT1mapping method accounts for fat and confounders through simultaneous estimation ofT1, , PDFF and . It demonstrates strong linear agreement with referenceT1measurements, with low bias and high precision, and can achieve full liver coverage in a single breath‐hold.
Purpose: Quantitative T1 mapping has the potential to replace biopsy for noninvasive diagnosis and quantitative staging of chronic liver disease. Conventional T1 mapping methods are confounded by fat and inhomogeneities, resulting in unreliable T1 estimations. Furthermore, these methods trade off spatial resolution and volumetric coverage for shorter acquisitions with only a few images obtained within a breath‐hold. This work proposes a novel, volumetric (3D), free‐breathing T1 mapping method to account for multiple confounding factors in a single acquisition.Theory and Methods: Free‐breathing, confounder‐corrected T1 mapping was achieved through the combination of non‐Cartesian imaging, magnetization preparation, chemical shift encoding, and a variable flip angle acquisition. A subspace‐constrained, locally low‐rank image reconstruction algorithm was employed for image reconstruction. The accuracy of the proposed method was evaluated through numerical simulations and phantom experiments with a T1/proton density fat fraction phantom at 3.0 T. Further, the feasibility of the proposed method was investigated through contrast‐enhanced imaging in healthy volunteers, also at 3.0 T.Results: The method showed excellent agreement with reference measurements in phantoms across a wide range of T1 values (200 to 1000 ms, slope = 0.998 (95% confidence interval (CI) [0.963 to 1.035]), intercept = 27.1 ms (95% CI [0.4 54.6]), r2 = 0.996), and a high level of repeatability. In vivo imaging studies demonstrated moderate agreement (slope = 1.099 (95% CI [1.067 to 1.132]), intercept = −96.3 ms (95% CI [−82.1 to −110.5]), r2 = 0.981) compared to saturation recovery‐based T1 maps.Conclusion: The proposed method produces whole‐liver, confounder‐corrected T1 maps through simultaneous estimation of T1, proton density fat fraction, and in a single, free‐breathing acquisition and has excellent agreement with reference measurements in phantoms.
Background: Breath‐held quantitative three‐dimensional chemical‐shift‐encoded (3D‐CSE)‐MRI is widely used for liver fat and iron quantification but is limited in patients unable to hold breath. Motion‐insensitive two‐dimensional flip angle modulated (2D‐FAM) acquisitions with short temporal aperture may address this limitation.Purpose: To evaluate the performance of 2D‐FAM‐based CSE‐MRI during free‐breathing for quantifying proton density fat‐fraction (PDFF) and R2*‐based liver iron concentration (LIC).Study Type: Retrospective single‐center observational analysis.Population: 230 patients (110 women [47.8%], mean age 51 ± 16 years [standard deviation, SD], range 4–85), undergoing routine clinical care between January and August 2024.Field Strength/Sequence: 1.5‐T and 3.0‐T, comparison of 2D‐FAM‐CSE‐MRI under free‐breathing and breath‐held 3D‐CSE‐MRI.Assessment: Three radiologists evaluated image quality, apparent signal‐to‐noise ratio (SNR), and motion artifacts on a 5‐point scale. PDFF and R2*‐based LIC were measured in each liver segment from each acquisition. Agreement of PDFF and LIC between techniques was assessed, comparing body mass index (BMI) groups (< 30 and ≥ 30 kg/m2). Inter‐rater reliability was assessed for all analyses.Statistical Tests: Inter‐rater reliability with intra‐class coefficient (ICC). Bias and limits of agreement (LoA) using Bland–Altman analysis. Linear regression assessed correlations of PDFF and LIC. Receiver operating characteristic (ROC) curves (AUC) to assess diagnostic performance. Statistical significance was set at p < 0.05 (Bonferroni‐corrected p < 0.01 for patient characteristics).Results: 2D‐FAM demonstrated superior image quality (median rating 4 vs. 3) and fewer motion artifacts (median rating 4 vs. 3) than 3D‐CSE‐MRI in PDFF and R2* maps. Inter‐rater reliability was good to excellent (ICC 0.77–0.92). PDFF agreement was excellent (bias: +0.32%, LoA [−1.37,2.00] for BMI < 30 kg/m2; +0.66%, [−2.44,3.76] for BMI ≥ 30 kg/m2). LIC agreement was high (bias: −0.02 mg/g, [−0.22,0.18]) up to a threshold of 7.0 mg/g. ROC analysis showed AUC up to 1.00 for PDFF and LIC thresholds.Data Conclusion: 2D‐FAM CSE‐MRI enables motion‐insensitive liver fat and iron quantification that can be performed during free‐breathing.Evidence Level: 2.Technical Efficacy: Stage 2.
2 September 2025 in Magnetic Resonance in Medicine
Purpose: To implement, optimize, and validate parallel imaging (PI)‐accelerated, 2D, flip angle modulated (FAM) chemical shift‐encoded quantification of liver proton‐density fat fraction (PDFF), with motion insensitivity. Methods: The optimization cost function that determines flip angles in FAM was generalized for PI. Phantom studies and prospective studies in volunteers with varying liver fat levels were performed. Free‐breathing FAM was acquired in the axial, sagittal, and coronal planes, with varying nominal PI acceleration factors ( R ) of 1.0 to 3.0. A breath‐held, commercially available 3D chemical shift‐encoded method was acquired as reference for PDFF. Overall image quality, qualitative SNR, and motion artifacts for all methods were Likert‐scale rated. PDFF measured by FAM was compared to reference to assess bias. Test–retest repeatability was assessed for all methods by repeating acquisitions after volunteer repositioning. Noise performance was assessed with standard deviation of PDFF maps as R increased. Results: The reader study ( N = 3 readers/10 subjects) demonstrated excellent image quality for FAM during free‐breathing, with reduced motion artifacts compared to breath‐held reference ( p < 0.01). PI‐accelerated FAM shows fewer motion artifacts than unaccelerated FAM ( p < 0.01). In all planes and accelerations, PDFF measured by FAM showed good agreement with reference PDFF measurements (mean bias: −0.4% to 2.0% PDFF; 95% limits of agreement: 2.8% to 4.0% PDFF). FAM in axial and coronal planes showed similar or improved repeatability (repeatability coefficient = 1.7% to 2.6% PDFF) compared to the reference (2.7%). Sagittal FAM shows similar or worse repeatability (repeatability coefficient = 3.0% to 3.6%). FAM with R = 2.0 has good noise performance and high SNR efficiency. Conclusion: FAM, in axial or coronal planes with R = 2.0, is optimal for motion‐insensitive liver PDFF quantification.