Leonardo Bianchi

Dr. Leonardo Bianchi

Politecnico di Milano
orcid

Research Focus & Metrics

Academic Career Length

7Years (2020-Present)

(Calculated once Scilit publication data is synced: current year − year of first publication + 1)

Representative Publications

View All Publications
6 February 2022 in International Journal of Hyperthermia
Verified
Journal Article
Open access
Leonardo BianchiLeonardo Bianchi
Leonardo Bianchi
FCFabiana Cavarzan
Fabiana Cavarzan
LCLucia Ciampitti
Lucia Ciampitti
MCMatteo Cremonesi
Matteo Cremonesi
FGFrancesca Grilli
Francesca Grilli
Paola SaccomandiPaola Saccomandi
Paola Saccomandi
Detailed information on the temperature dependence of tissue thermophysical and mechanical properties is pivotal for the optimal implementation of mathematical models and simulation-based tools for...

21 June 2021 in Sensors
Verified
Journal Article
Open access
PDF file
Ahad MohammadiAhad Mohammadi
Ahad Mohammadi
Leonardo BianchiLeonardo Bianchi
Leonardo Bianchi
Somayeh AsadiSomayeh Asadi
Somayeh Asadi
Paola SaccomandiPaola Saccomandi
Paola Saccomandi
The ability to predict heat transfer during hyperthermal and ablative techniques for cancer treatment relies on understanding the thermal properties of biological tissue. In this work, the thermal properties of ex vivo liver, pancreas and brain tissues are reported as a function of temperature. The thermal diffusivity, thermal conductivity and volumetric heat capacity of these tissues were measured in the temperature range from 22 to around 97 °C. Concerning the pancreas, a phase change occurred around 45 °C; therefore, its thermal properties were investigated only until this temperature. Results indicate that the thermal properties of the liver and brain have a non-linear relationship with temperature in the investigated range. In these tissues, the thermal properties were almost constant until 60 to 70 °C and then gradually changed until 92 °C. In particular, the thermal conductivity increased by 100% for the brain and 60% for the liver up to 92 °C, while thermal diffusivity increased by 90% and 40%, respectively. However, the heat capacity did not significantly change in this temperature range. The thermal conductivity and thermal diffusivity were dramatically increased from 92 to 97 °C, which seems to be due to water vaporization and state transition in the tissues. Moreover, the measurement uncertainty, determined at each temperature, increased after 92 °C. In the temperature range of 22 to 45 °C, the thermal properties of pancreatic tissue did not change significantly, in accordance with the results for the brain and liver. For the three tissues, the best fit curves are provided with regression analysis based on measured data to predict the tissue thermal behavior. These curves describe the temperature dependency of tissue thermal properties in a temperature range relevant for hyperthermia and ablation treatments and may help in constructing more accurate models of bioheat transfer for optimization and pre-planning of thermal procedures.

21 April 2021 in Biomedical Optics Express
Verified
Journal Article
Closed access
Leonardo BianchiLeonardo Bianchi
Leonardo Bianchi
Sanzhar KorganbayevSanzhar Korganbayev
Sanzhar Korganbayev
Annalisa OrricoAnnalisa Orrico
Annalisa Orrico
MLMartina De Landro
Martina De Landro
Paola SaccomandiPaola Saccomandi
Paola Saccomandi
This work proposes the quasi-distributed real-time monitoring and control of laser ablation (LA) of liver tissue. To confine the thermal damage, a pre-planning stage of the control strategy based on numerical simulations of the bioheat-transfer was developed to design the control parameters, then experimentally assessed. Fiber Bragg grating (FBG) sensors were employed to design the automatic thermometry system used for temperature feedback control for interstitial LA. The tissue temperature was maintained at a pre-set value, and the influence of different sensor locations (on the direction of the beam propagation and backward) on the thermal outcome was evaluated in comparison with the uncontrolled case. Results show that the implemented computational model was able to properly describe the temperature evolution of the irradiated tissue. Furthermore, the realized control strategy allowed for the accurate confinement of the laser-induced temperature increase, especially when the temperature control was actuated by sensors located in the direction of the beam propagation, as confirmed by the calculated fractions of necrotic tissues (e.g., 23 mm3 and 53 mm3 for the controlled and uncontrolled LA, respectively).

6 August 2020 in Journal of Biophotonics
Verified
Journal Article
Open access
PDF file
Somayeh AsadiSomayeh Asadi
Somayeh Asadi
Leonardo BianchiLeonardo Bianchi
Leonardo Bianchi
MLMartina De Landro
Martina De Landro
SKSanzhar Korganbayev
Sanzhar Korganbayev
ESEmiliano Schena
Emiliano Schena
Paola SaccomandiPaola Saccomandi
Paola Saccomandi
Gold nanoparticles (GNPs)‐based photothermal therapy (PTT) is a promising minimally invasive thermal therapy for the treatment of focal malignancies. Although GNPs‐based PTT has been known for over two decades and GNPs possess unique properties as therapeutic agents, the delivery of a safe and effective therapy is still an open question. This review aims at providing relevant and recent information on the usage of GNPs in combination with the laser to treat cancers, pointing out the practical aspects that bear on the therapy outcome. Emphasis is given to the assessment of the GNPs’ properties and the physical mechanisms underlying the laser‐induced heat generation in GNPs‐loaded tissues. The main techniques available for temperature measurement and the current theoretical simulation approaches predicting the therapeutic outcome are reviewed. Topical challenges in delivering safe thermal dosage are also presented with the aim to discuss the state‐of‐the‐art and the future perspective in the field of GNPs‐mediated PTT.

13 November 2020 in Sensors
Verified
Journal Article
Open access
PDF file
Sanzhar KorganbayevSanzhar Korganbayev
Sanzhar Korganbayev
Annalisa OrricoAnnalisa Orrico
Annalisa Orrico
Leonardo BianchiLeonardo Bianchi
Leonardo Bianchi
MLMartina De Landro
Martina De Landro
AWAlexey Wolf
Alexey Wolf
Paola SaccomandiPaola Saccomandi
Paola Saccomandi
Laser ablation (LA) of cancer is a minimally invasive technique based on targeted heat release. Controlling tissue temperature during LA is crucial to achieve the desired therapeutic effect in the organs while preserving the healthy tissue around. Here, we report the design and implementation of a real-time monitoring system performing closed-loop temperature control, based on fiber Bragg grating (FBG) spatial measurements. Highly dense FBG arrays (1.19 mm length, 0.01 mm edge-to-edge distance) were inscribed in polyimide-coated fibers using the femtosecond point-by-point writing technology to obtain the spatial resolution needed for accurate reconstruction of high-gradient temperature profiles during LA. The zone control strategy was implemented such that the temperature in the laser-irradiated area was maintained at specific set values (43 and 55 °C), in correspondence to specific radii (2 and 6 mm) of the targeted zone. The developed control system was assessed in terms of measured temperature maps during an ex vivo liver LA. Results suggest that the temperature-feedback system provides several advantages, including controlling the margins of the ablated zone and keeping the maximum temperature below the critical values. Our strategy and resulting analysis go beyond the state-of-the-art LA regulation techniques, encouraging further investigation in the identification of the optimal control-loop.

20 September 2021 in IEEE Transactions on Instrumentation and Measurement
Verified
Journal Article
Open access
Sanzhar KorganbayevSanzhar Korganbayev
Sanzhar Korganbayev
Annalisa OrricoAnnalisa Orrico
Annalisa Orrico
Leonardo BianchiLeonardo Bianchi
Leonardo Bianchi
DPDavide Paloschi
Davide Paloschi
AWAlexey Wolf
Alexey Wolf
Paola SaccomandiPaola Saccomandi
Paola Saccomandi
In this article, we propose a temperature-based proportional–integral–derivative (PID) controlling algorithm using highly dense fiber Bragg grating (FBG) arrays for laser ablation (LA) of ex vivo pancreatic tissues. Custom-made highly dense FBG arrays with a spatial resolution of 1.2 mm were fabricated with the femtosecond point-by-point writing technology and optimized for LA applications. In order to obtain proper PID gain values, finite element method-based iterative simulation of different PID gains was performed. Then, the proposed algorithm, with numerically derived PID gains, was experimentally validated. In the experiments, the point temperature was controlled at different distances from the laser fiber tip (6.0, 7.2, 8.4, and 10.8 mm). The obtained results report robust controlling and correlation between controlled distance and the resulting area of ablation. The results of the work encourage further investigation of FBG array application for LA control.