21 April 2021 in Biomedical Optics Express
Verified
Journal Article
Closed access
Leonardo BianchiLeonardo Bianchi
Leonardo Bianchi
Sanzhar KorganbayevSanzhar Korganbayev
Sanzhar Korganbayev
Annalisa OrricoAnnalisa Orrico
Annalisa Orrico
MLMartina De Landro
Martina De Landro
Paola SaccomandiPaola Saccomandi
Paola Saccomandi
This work proposes the quasi-distributed real-time monitoring and control of laser ablation (LA) of liver tissue. To confine the thermal damage, a pre-planning stage of the control strategy based on numerical simulations of the bioheat-transfer was developed to design the control parameters, then experimentally assessed. Fiber Bragg grating (FBG) sensors were employed to design the automatic thermometry system used for temperature feedback control for interstitial LA. The tissue temperature was maintained at a pre-set value, and the influence of different sensor locations (on the direction of the beam propagation and backward) on the thermal outcome was evaluated in comparison with the uncontrolled case. Results show that the implemented computational model was able to properly describe the temperature evolution of the irradiated tissue. Furthermore, the realized control strategy allowed for the accurate confinement of the laser-induced temperature increase, especially when the temperature control was actuated by sensors located in the direction of the beam propagation, as confirmed by the calculated fractions of necrotic tissues (e.g., 23 mm3 and 53 mm3 for the controlled and uncontrolled LA, respectively).

13 November 2020 in Sensors
Verified
Journal Article
Open access
PDF file
Sanzhar KorganbayevSanzhar Korganbayev
Sanzhar Korganbayev
Annalisa OrricoAnnalisa Orrico
Annalisa Orrico
Leonardo BianchiLeonardo Bianchi
Leonardo Bianchi
MLMartina De Landro
Martina De Landro
AWAlexey Wolf
Alexey Wolf
Paola SaccomandiPaola Saccomandi
Paola Saccomandi
Laser ablation (LA) of cancer is a minimally invasive technique based on targeted heat release. Controlling tissue temperature during LA is crucial to achieve the desired therapeutic effect in the organs while preserving the healthy tissue around. Here, we report the design and implementation of a real-time monitoring system performing closed-loop temperature control, based on fiber Bragg grating (FBG) spatial measurements. Highly dense FBG arrays (1.19 mm length, 0.01 mm edge-to-edge distance) were inscribed in polyimide-coated fibers using the femtosecond point-by-point writing technology to obtain the spatial resolution needed for accurate reconstruction of high-gradient temperature profiles during LA. The zone control strategy was implemented such that the temperature in the laser-irradiated area was maintained at specific set values (43 and 55 °C), in correspondence to specific radii (2 and 6 mm) of the targeted zone. The developed control system was assessed in terms of measured temperature maps during an ex vivo liver LA. Results suggest that the temperature-feedback system provides several advantages, including controlling the margins of the ablated zone and keeping the maximum temperature below the critical values. Our strategy and resulting analysis go beyond the state-of-the-art LA regulation techniques, encouraging further investigation in the identification of the optimal control-loop.

20 September 2021 in IEEE Transactions on Instrumentation and Measurement
Verified
Journal Article
Open access
Sanzhar KorganbayevSanzhar Korganbayev
Sanzhar Korganbayev
Annalisa OrricoAnnalisa Orrico
Annalisa Orrico
Leonardo BianchiLeonardo Bianchi
Leonardo Bianchi
DPDavide Paloschi
Davide Paloschi
AWAlexey Wolf
Alexey Wolf
Paola SaccomandiPaola Saccomandi
Paola Saccomandi
In this article, we propose a temperature-based proportional–integral–derivative (PID) controlling algorithm using highly dense fiber Bragg grating (FBG) arrays for laser ablation (LA) of ex vivo pancreatic tissues. Custom-made highly dense FBG arrays with a spatial resolution of 1.2 mm were fabricated with the femtosecond point-by-point writing technology and optimized for LA applications. In order to obtain proper PID gain values, finite element method-based iterative simulation of different PID gains was performed. Then, the proposed algorithm, with numerically derived PID gains, was experimentally validated. In the experiments, the point temperature was controlled at different distances from the laser fiber tip (6.0, 7.2, 8.4, and 10.8 mm). The obtained results report robust controlling and correlation between controlled distance and the resulting area of ablation. The results of the work encourage further investigation of FBG array application for LA control.

19 October 2021 in Sensors
Verified
Journal Article
Open access
PDF file
MLMartina De Landro
Martina De Landro
Eric FelliEric Felli
Eric Felli
TCToby Collins
Toby Collins
RNRichard Nkusi
Richard Nkusi
ABAndrea Baiocchini
Andrea Baiocchini
Paola SaccomandiPaola Saccomandi
Paola Saccomandi
Thermal ablation is an acceptable alternative treatment for primary liver cancer, of which laser ablation (LA) is one of the least invasive approaches, especially for tumors in high-risk locations. Precise control of the LA effect is required to safely destroy the tumor. Although temperature imaging techniques provide an indirect measurement of the thermal damage, a degree of uncertainty remains about the treatment effect. Optical techniques are currently emerging as tools to directly assess tissue thermal damage. Among them, hyperspectral imaging (HSI) has shown promising results in image-guided surgery and in the thermal ablation field. The highly informative data provided by HSI, associated with deep learning, enable the implementation of non-invasive prediction models to be used intraoperatively. Here we show a novel paradigm “peak temperature prediction model” (PTPM), convolutional neural network (CNN)-based, trained with HSI and infrared imaging to predict LA-induced damage in the liver. The PTPM demonstrated an optimal agreement with tissue damage classification providing a consistent threshold (50.6 ± 1.5 °C) for the damage margins with high accuracy (~0.90). The high correlation with the histology score (r = 0.9085) and the comparison with the measured peak temperature confirmed that PTPM preserves temperature information accordingly with the histopathological assessment.

1 June 2020
Verified
Conference Paper
Closed access
SKSanzhar Korganbayev
Sanzhar Korganbayev
RPRiccardo Pini
Riccardo Pini
Annalisa OrricoAnnalisa Orrico
Annalisa Orrico
AWAlexey Wolf
Alexey Wolf
ADAlexander Dostovalov
Alexander Dostovalov
Paola SaccomandiPaola Saccomandi
Paola Saccomandi
In this paper we present a temperature-controlled laser ablation technique based on fiber Bragg grating (FBG) array temperature measurements. Highly dense FBG array with 0.9 mm-long gratings was specially fabricated with the femtosecond point-by-point writing technology to have the high spatial resolution needed for accurate temperature profile measurements during laser ablation. Real-time temperature monitoring was used for laser power regulation based on ON-OFF control logic to maintain stable peak temperature during ablation. Two temperature thresholds and two comparison times $\Delta \tau$ were set: 43 and 48 °C, and 1 and 2.5 s, respectively. The outcomes of these settings were evaluated in terms of temperature distribution in an ex vivo porcine liver undergoing laser ablation. Results show that the proposed use of FBG array for controlled ablation provides several advantages over regulation with traditional thermometers. Quasi-distributed sensing capability of FBG array allows obtaining information about spatial heat distribution during ablation. Moreover, this study points out the importance of considering not only temperature but also the effect of $\Delta \tau$ on both tissue outcome and laser system safety. When $\Delta \tau$ =2.5 s the control action is less aggressive and temperature spikes are more observable, whereas for $\Delta \tau$ =1 s the set point is reached faster and is better followed. Additionally, smaller $\Delta\tau$ leads to smooth temperature control and a spatially confined ablation region. The results of this study encourage further investigation of the optimal control-loop and laser system settings for improving LA effects while avoiding carbonization, preserving the light penetration and obtaining reproducible outcomes.