We examine the broadband behavior of complex electrical properties of glycerin and water mixtures over the frequency range of 0.1-25.0 GHz, especially as they relate to using these liquids as coupling media for microwave tomographic imaging. Their combination is unique in that they are mutually miscible over the full range of concentrations, which allows them to be tailored to dielectric property matching for biological tissues. While the resultant mixture properties are partially driven by differences in the inherent low-frequency permittivity of each constituent, relaxation frequency shifts play a disproportionately larger role in increasing the permittivity dispersion while also dramatically increasing the effective conductivity over the frequency range of 1-3 GHz. For the full range of mixture ratios, the relaxation frequency shifts from 17.5 GHz for 0% glycerin to less than 0.1 GHz for 100% glycerin. Of particular interest is the fact that the conductivity stays above 1.0 S/m over the 1-3-GHz range for glycerin mixture ratios (70%-90% glycerin) we use for microwave breast tomography. The high level of attenuation is critical for suppressing unwanted multipath signals. This paper presents a full characterization of these liquids along with a discussion of their benefits and limitations in the context of microwave tomography.
Introduction: Microwave tomography recovers images of tissue dielectric properties, which appear to be specific for breast cancer, with low-cost technology that does not present an exposure risk, suggesting the modality may be a good candidate for monitoring neoadjuvant chemotherapy. Methods: Eight patients undergoing neoadjuvant chemotherapy for locally advanced breast cancer were imaged longitudinally five to eight times during the course of treatment. At the start of therapy, regions of interest (ROIs) were identified from contrast-enhanced magnetic resonance imaging studies. During subsequent microwave examinations, subjects were positioned with their breasts pendant in a coupling fluid and surrounded by an immersed antenna array. Microwave property values were extracted from the ROIs through an automated procedure and statistical analyses were performed to assess short term (30 days) and longer term (four to six months) dielectric property changes. Results: Two patient cases (one complete and one partial response) are presented in detail and demonstrate changes in microwave properties commensurate with the degree of treatment response observed pathologically. Normalized mean conductivity in ROIs from patients with complete pathological responses was significantly different from that of partial responders (P value = 0.004). In addition, the normalized conductivity measure also correlated well with complete pathological response at 30 days (P value = 0.002). Conclusions: These preliminary findings suggest that both early and late conductivity property changes correlate well with overall treatment response to neoadjuvant therapy in locally advanced breast cancer. This result is consistent with earlier clinical outcomes that lesion conductivity is specific to differentiating breast cancer from benign lesions and normal tissue.
10 January 2019 in IEEE Transactions on Biomedical Engineering
Objective: Fusion of magnetic resonance imaging (MRI) breast images with microwave tomography is accomplished through a soft prior technique, which incorporates spatial information (from MRI), i.e., accurate boundary location of different regions of interest, into the regularization process of the microwave image reconstruction algorithm. Methods: Numerical experiments were completed on a set of three-dimensional (3-D) breast geometries derived from MR breast data with different parenchymal densities, as well as a simulated tumor to evaluate the performance over a range of breast shapes, sizes, and property distributions. Results: When the soft prior regularization technique was applied, both permittivity and conductivity relative root mean square error values decreased by more than 87% across all breast densities, except in two cases where the error decrease was only 55% and 78%. In addition, the incorporation of structural priors increased contrast between tumor and fibroglandular tissue by 59% in permittivity and 192% in conductivity. Conclusion:This study confirmed that the soft prior algorithm is robust in 3-D and can function successfully across a range of complex geometries and tissue property distributions. Significance: This study demonstrates that our microwave tomography is capable of recovering accurate tissue property distributions when spatial information from MRI is incorporated through soft prior regularization.
A breast phantom developed at the Supelec Institute was interrogated to study its suitability for microwave tomography measurements. A microwave measurement system based on 16 monopole antennas and a vector network analyzer was used to study how the S-parameters are influenced by insertion of the phantom. The phantom is a 3D-printed structure consisting of plastic shells that can be filled with tissue mimicking liquids. The phantom was filled with different liquids and tested with the measurement system to determine whether the plastic has any effects on the recovered images or not. Measurements of the phantom when it is filled with the same liquid as the surrounding coupling medium are of particular interest. In this case, the phantom plastic has a substantial effects on the measurements which ultimately detracts from the desired images.
5 March 2021 in IEEE Transactions on Microwave Theory and Techniques
This article describes a fast microwave tomography reconstruction algorithm based on the 2-D discrete dipole approximation (DDA). Synthetic data from a finite-element-based solver and experimental data from a microwave imaging system are used to reconstruct images and to validate the algorithm. The microwave measurement system consists of 16 monopole antennas immersed in a tank filled with lossy coupling liquid and a vector network analyzer. The low-profile antennas and lossy nature of the system make the DDA an ideal forward solver in image reconstructions. The results show that the algorithm can readily reconstruct a 2-D plane of a cylindrical phantom. The proposed forward solver combined with the nodal adjoint method for computing the Jacobian matrix enables the algorithm to reconstruct an image within 6 s. This implementation provides significant time savings and reduced memory requirements and is a dramatic improvement over previous implementations.
We have developed a multichannel software defined radio-based transceiver measurement system for use in general microwave tomographic applications. The unit is compact enough to fit conveniently underneath the current illumination tank of the Dartmouth microwave breast imaging system. The system includes 16 channels that can both transmit and receive and it operates from 500 MHz to 2.5 GHz while measuring signals down to −140 dBm. As is the case with multichannel systems, cross-channel leakage is an important specification and must be lower than the noise floors for each receiver. This design exploits the isolation inherent when the individual receivers for each channel are physically separate; however, these challenging specifications require more involved signal isolation techniques at both the system design level and the individual, shielded component level. We describe the isolation design techniques for the critical system elements and demonstrate specification compliance at both the component and system level.