Neural prostheses are limited by the availability of peripheral neural electrodes to record the user's intention or provide sensory feedback through functional electrical stimulation. Our objective was to compare the ability of the novel “transverse intrafascicular multi-channel electrode” (TIME) and an earlier generation “thin-film longitudinal intrafascicular electrode” (tfLIFE) to selectively stimulate nerve fascicles and activate forelimb muscles in pigs. TIME was designed to access a larger subpopulation of fascicles than tfLIFE and should therefore be able to selectively activate a larger number of muscles. Electrodes were implanted in the median nerve, and sequential electric stimulation was applied to individual contacts. The compound muscle action potentials of seven muscles were recorded to quantify muscle recruitment. As expected, TIME was able to recruit more muscles with higher selectivity than tfLIFE (significant difference when comparing the performance of an entire electrode); a similar activation current was used (no significant difference). Histological analysis revealed that electrodes were located between fascicles, which influenced the selectivity and activation current level. In conclusion, TIME is a viable neural interface for selective activation of multiple fascicles in human-sized nerves that may assist to pave the way for future neuroprosthesis applications.
After an incomplete spinal cord injury (SCI), we know that plastic reorganization occurs in supraspinal structures with residual descending tracts. However, our knowledge about spinal plasticity is rather limited. Our recent studies point to changes within the spinal cord below the lesion. After a lateral left hemisection (T10), cats recovered stepping with both hindlimbs within 3 weeks. After a complete section (T13) in these cats, bilateral stepping was seen on the next day, a skill usually acquired after several weeks of treadmill training. This indicates that durable plastic changes occurred below the lesion. However, because sensory feedback entrains the stepping rhythm, it is difficult to reveal central pattern generator (CPG) adaptation. Here, we investigated whether lumbar segments of cats with a chronic hemisection were able to generate fictive locomotion—that is, without phasic sensory feedback as monitored by five muscle nerves in each hindlimb. With a chronic left hemisection, the number of muscle nerves displaying locomotor bursts was larger on the left than on the right. In addition, transmission of cutaneous reflexes was relatively facilitated on the left. Later during the acute experiment, a complete spinalization (T13) was performed and clonidine was injected to induce rhythmic activities. There were still more muscle nerves displaying locomotor bursts on the left. The results demonstrate that spinal networks were indeed modified after a hemisection with a clear asymmetry between left and right in the capacity to generate locomotion. Plastic changes in CPG and reflex transmission below the lesion are thus involved in the stepping recovery after an incomplete SCI.
Over the last three years, a novel type of intrafascicular electrode was developed to serve as a corrective neural prosthetic interface to be placed in the peripheral nerve stumps of upper limb amputees who suffer from phantom limb pain (PLP). In general, the presented interface is an advancement of the thin-film LIFE electrode that was successfully used in amputees a couple of years ago. Although the LIFE concept proved to be effective, the longitudinal orientation of the implant prevented distribution of spatial selective sites in more than one fascicle. Therefore, a transversal approach was developed. The work presented was developed within the European project “TIME” and will be implanted in the median and ulnar nerves of an amputee subject in Rome later this year (2012). The present paper will outline the design, assembly technologies and full electrochemical characterization of the TIME-3H electrode for sub-chronic (<;30 days) human implants.
The adoption of high-density electrode systems for human–machine interfaces in real-life applications has been impeded by practical and technical challenges, including noise interference, motion artefacts and the lack of compact electrode interfaces. To overcome some of these challenges, we introduce a wearable and stretchable electromyography (EMG) array, and present its design, fabrication methodology, characterisation, and comprehensive evaluation. Our proposed solution comprises dry-electrodes on flexible printed circuit board (PCB) substrates, eliminating the need for time-consuming skin preparation. The proposed fabrication method allows the manufacturing of stretchable sleeves, with consistent and standardised coverage across subjects. We thoroughly tested our developed prototype, evaluating its potential for application in both research and real-world environments. The results of our study showed that the developed stretchable array matches or outperforms traditional EMG grids and holds promise in furthering the real-world translation of high-density EMG for human–machine interfaces.
16 December 2019 in Journal of Micromechanics and Microengineering
In this work, the development of an active high-density transverse intrafascicular micro-electrode (hd-TIME) probe to interface with the peripheral nervous system is presented. The TIME approach is combined with an active probe chip, resulting in improved selectivity and excellent signal-to-noise ratio. The integrated multiplexing capabilities reduce the number of external electrical connections and facilitate the positioning of the probe during implantation, as the most interesting electrodes of the electrode array can be selected after implantation. The probe chip is packaged using thin-film manufacturing techniques to allow for a minimally invasive electronic package. Special attention is paid to the miniaturization, the mechanical flexibility and the hermetic encapsulation of the device. A customized probe chip was designed and packaged using a flexible, implantable thin electronic package (FITEP) process platform. The platform is specifically developed for making slim, ultra-compliant, implantable complementary metal-oxide-semiconductor based electronic devices. Multilayer stacks of polyimide films and HfO 2 /Al 2 O 3 /HfO 2 layers deposited via atomic layer deposition act as bidirectional diffusion barriers and are key to the hermetic encapsulation. Their efficacy was demonstrated both by water vapor transmission rate tests and accelerated immersion tests in phosphate buffered saline at 60 °C. Using the hd-TIME probe, an innovative implantation method is developed to prevent the fascicles from moving away when the epineurium is pierced. In addition, by transversally implanting the hd-TIME probe in the proximal sciatic nerve of a rat, selective activation within the nerve was demonstrated. The FITEP process platform can be applied to a broader range of integrated circuits and can be considered as an enabler for other biomedical applications.