For the last 20 years, researchers have developed accelerometers to function as ossicular vibration sensors in order to eliminate the external components of hearing aid and cochlear implant (CI) systems. To date, no accelerometer has met all of the stringent performance requirements necessary to function in this capacity. In this work, we present an accelerometer design with an equivalent noise floor of less than 20 phon equal-loudness-level over a 0.1–8-kHz bandwidth in a package small enough to be implanted in the middle ear. Our approach uses a dual-bandwidth (two sensing elements) microelectromechanical systems (MEMS) piezoelectric accelerometer, sized using an area-minimization process based on an experimentally validated analytical model of the sensor. The resulting bandwidth of the low-frequency sensing element (LFSE) is 0.1–1.25 kHz and that of the high-frequency sensing element (HFSE) is 1.25–8 kHz. These sensing elements fit within a silicon frame that is $795\times 778\,\,\mu \text{m}$ , which can reasonably be housed along with a required integrated circuit (IC) in a $2.2\times 2.7\times1$ -mm package. The estimated total mass of the packaged system is approximately 14 mg. This dual-bandwidth MEMS sensor fills a technological gap in current completely implantable auditory prosthesis (CIAP) research and development by enabling a device capable of meeting physical and performance specifications needed for use in the middle ear.
In this study, a method for determining the optimal location and orientation of an implantable piezoelectric accelerometer on the short process of the incus is presented. The accelerometer is intended to be used as a replacement for an external microphone to enable totally implantable auditory prostheses. The optimal orientation of the sensor and the best attachment point are determined based on two criteria—maximum pressure sensitivity sum and minimum loudness level sum. The best location is determined to be near the incudomalleolar joint. We find that the angular orientation of the sensor is critical and provide guidelines on that orientation. The method described in this paper can be used to further optimize the design and performance of the accelerometer.
1 June 2025 in The Journal of the Acoustical Society of America
Hearing aids and cochlear implants help patients treat their hearing loss, but have limitations impacting their use rates. Completely implantable auditory prostheses would expand the range of activities a prosthesis user could engage in and enable 24/7 use. However, the lack of a completely implantable microphone that is robust, lightweight, and low noise prevents the wide adoption of implantable devices. Current implantable sensors struggle to meet or exceed the performance necessary for this application. This work develops a discretized and exhaustive design optimization approach to identify multi-bandwidth transducers that meet the 20-phon noise floor over 100 Hz–8 kHz. The design procedure is based on an experimentally validated analytical model that simulates the response of miniature piezoelectric microelectromechanical systems (MEMS) accelerometers. A four-bandwidth accelerometer with constrained proof mass thicknesses is selected as the design that best balances area minimization with ease of manufacturability. The estimated MEMS die dimensions are 825 μm × 575 μm, which is a 23% MEMS die area reduction compared to the previously published dual-bandwidth sensor [A. E. Hake, P. Kitsopoulos, and K. Grosh, IEEE Sens. J., 23(13), 13957–13965 (2023)].
1 March 2023 in The Journal of the Acoustical Society of America
Moderate to severe hearing loss is a debilitating condition that affects over 5% of the world’s population. Hearing aids and cochlear implants positively impact the lives of those who suffer from sensorineural hearing loss. However, both display numerous limitations that affect their adoption and use rates, notably including those associated with the external elements of these devices (e.g., microphone and signal processor). These external components impact device safety, appearance, acoustic performance, and ease-of-use. A totally implantable auditory prosthesis would help to address these issues by eliminating external components. A major barrier to progress toward this goal is the lack of a completely implantable acoustic sensor capable of matching or exceeding the performance of commercial external microphones. Our previous studies have indicated that piezoelectric microelectromechanical systems (MEMS) accelerometers have the potential to function as implantable sensors within the middle ear meeting a 20-phon noise floor over a 100–8 kHz range. In this paper, we describe a process to design and fabricate dual and tri-resonance devices that are comprised of piezoelectric cantilever bimorph beams tip-loaded by a proof mass, can meet or exceed the 20-phon noise floor, and can produce a measurable voltage output when deflected by sound-induced ossicular vibrations.