The purpose of this study was to measure out-of-field organ doses in clinical conditions in anthropomorphic paediatric phantoms which received a simulated treatment of a brain tumour with intensity modulated radiotherapy (IMRT) and 3D conformal radiotherapy (3D CRT). Organ doses measured with radiophotoluminescent and thermoluminescent dosemeters were on average 1.6 and 3.0 times higher for the 5 y-old than for the 10 y-old phantom for IMRT and 3D CRT, respectively. A larger 5-y to 10-y organ dose ratio for 3D CRT can be explained because the use of a mechanical wedge for the 5-y-old 3D CRT phantom treatment increased out-of-field doses. Due to different configurations of the radiation fields, for both phantoms, the IMRT technique resulted in a higher non-target brain dose and higher eye doses but lower thyroid doses compared to 3D CRT. For 3D CRT (which used a non-coplanar field configuration), eye doses were 3-6% and for IMRT (which used a coplanar field configuration) 27-30% of the treatment dose, respectively. For thyroid and more distant organs, doses were less than 1% of the treatment dose. Comparison of measured doses and doses calculated by the treatment planning system (TPS) showed that the TPS underestimated out-of-field doses both for IMRT and 3D CRT.
The purpose of this study was to measure out-of-field organ doses in two anthropomorphic child phantoms for the treatment of large brain arteriovenous malformations (AVMs) using hypofractionated gamma knife (GK) radiosurgery and to compare these with an alternative treatment using intensity-modulated radiation therapy (IMRT). Target volume was identical in size and shape in all cases. Radiophotoluminescent (RPL), thermoluminescent (TL) and optically stimulated luminescent (OSL) dosimeters were used for out-of-field dosimetry during GK treatment and a good agreement within 1–2% between results was shown. In addition, the use of multiple dosimetry systems strengthens the reliability of the findings. The number of GK isocentres was confirmed to be important for the magnitude of out-of-field doses. Measured GK doses for the same distance from the target, when expressed per target dose and isocentre, were comparable in both phantoms. GK out-of-field doses averaged for both phantoms were evaluated to be 120 mGy/Gy for eyes then sharply reduced to 20 mGy/Gy for mandible and slowly reduced up to 0.8 mGy/Gy for testes. Taking into account the fractionation regimen used to treat AVM patients, the total treatment organ doses to the out-of-field organs were calculated and compared with IMRT. The eyes were better spared with GK whilst for more distant organs doses were up to a factor of 2.8 and 4 times larger for GK compared to IMRT in 5-year and 10-year old phantoms, respectively. Presented out-of-field dose values are specific for the investigated AVM case, phantoms and treatment plans used for GK and IMRT, but provide useful information about out-of-field dose levels and emphasise their importance.