Background: The need to improve gait emulation in people with amputation has driven the development of customized prosthetic mechanisms. This study focuses on the design and validation of a mechanism for external knee joint prostheses, based on the trajectory of the Instantaneous Center of Rotation (ICR) of a healthy knee. Objective: The objective is to design a mechanism that accurately reproduces the evolution of the ICR trajectory, thereby improving stability and reducing the user’s muscular effort. Methods: An exploratory methodology was employed, utilizing computer-aided design (CAD), kinematic simulations, and rapid prototyping through 3D printing. Multiple configurations of four- and six-bar mechanisms were evaluated to determine the ICR trajectory and compare it with a reference model obtained in the laboratory from a specific subject, using MATLAB-2023a and the Fréchet distance as an error metric. Results: The results indicated that the four-bar mechanism, with the incorporation of a simple gear train, achieved a more accurate emulation of the ICR trajectory, reaching a minimum error of 6.87 mm. Functional tests confirmed the effectiveness of the design in terms of stability and voluntary control during gait. It can be concluded that integrating the mechanism with the gear train significantly enhances its functionality, making it a viable alternative for the development of external knee prostheses for people with transfemoral amputation, based on the ICR of the contralateral leg.
: The need to improve gait emulation in patients with transfemoral amputation has led to the development of customized prosthetic mechanisms. This study focuses on the design and validation of an external knee joint prosthesis mechanism, based on the trajectory of the Instantaneous Center of Rotation (ICR) of a healthy knee. The goal is to design a mechanism that closely follows the evolution of the ICR path, improving stability and reducing the user's muscle effort. An exploratory methodology was applied, using computer-aided design (CAD), kinematic simulations, and rapid prototyping with 3D printing. Several four-bar and six-bar mechanism configurations were evaluated by determining their ICR trajectories and comparing them to a reference model obtained in the lab from a specific patient, using MATLAB and Fréchet distance as the error metric. The results showed that the optimized four-bar mechanism, with the addition of gears, achieved the highest accuracy in matching the ICR trajectory, with a minimum error of 5 mm. Functional tests confirmed the effectiveness of the design in terms of stability and voluntary control during walking. It is concluded that optimizing the mechanism design based on the ICR significantly improves functionality and supports its clinical implementation and impact on users’ quality of life.