This research presents the results of laboratory nonlinear characterization of a diaphragm-bonded FBG sensor under coupled pressure–temperature loading. The experiments were conducted using a laboratory-scale pressure–temperature simulator over a pressure range of 0–2.0 MPa and a temperature range of 5–50°C, followed by combined pressure–temperature testing. The sensor’s response was found to be nonlinear. Polynomial models (quadratic or 3rd-order) consistently yielded better results than linear models. Local sensitivity varied with both pressure and temperature, ranging from 0.091 to 0.371 nm/MPa and from 0.003 to 0.022 nm/°C. Analysis using the Response Surface Method (RSM) showed that pressure contributes significantly to the slope of the sensitivity, while temperature affects the offset shift of the wavelength. Overall, the research results offer a practical method for characterizing the response of FBG sensors under controlled pressure and temperature conditions.
Fiber Bragg Grating (FBG) sensors are widely used to measure pressure and temperature. However, when both are subjected to a coupled environment, the Bragg wavelengths influence each other, causing biased measurements if not properly calibrated. Most calibration models assume linear sensitivity, which may not hold across the entire operating range. This study proposes a five-step statistical calibration framework for a single FBG sensor under coupled pressure–temperature conditions: data preparation; fitting linear, quadratic, and cubic models; model selection using AIC/BIC; local sensitivity calculation; and contribution decomposition via Response Surface Methodology (RSM). The framework was tested on experimental data from an FBG sensor mounted on a metal membrane, at pressures of 0 MPa–2.0 MPa and temperatures of 5 °C–45 °C. Nonlinear models outperformed linear ones, with values of 0.9725–0.9992. The model was validated using Leave-One-Out Cross-Validation (LOOCV), yielding an out-of-sample RMSE (Root Mean Square Error) of 0.0452 nm. Pressure sensitivity remained constant at 0.2250 nm/MPa, while temperature sensitivity decreased from 0.0224 nm/°C at 5 °C to 0.0069 nm/°C at 45 °C. RSM showed that pressure influences the slope and temperature influences the offset, with no significant interaction. Residual analysis satisfied the regression assumptions, and the combined uncertainty (0.008 nm/MPa; 0.0003 nm/°C) was much smaller than the observed sensitivity range, confirming that the variations are physical, not instrument noise. This framework is practical and statistically sound for single FBG sensor calibration.