This paper presents the model identification of a cryogenic current comparator (CCC). A dynamic model set is obtained and compared with the experimental data in order to provide a realistic dynamical behavior of the system. To improve the performance of the CCC, an H ∞ optimal controller is designed based on this model set. In this framework, a robust stability guarantee is provided and the simulations of the closed-loop system illustrate the performance and robustness improvements.
A pulse drive system capable of converting an arbitrary input signal into a delta sigma encoded pulse code stream in real time has been demonstrated. This system is based on the use of analog and digital hardware controlled by a Field Programmable Gate Array (FPGA) rather than using software to produce the delta sigma code and then loading it into the memory of a commercial Pulse Pattern Generator (PPG). The results of using the system to drive a Josephson junction array to synthesize quantum-accurate voltage waveforms are presented. The advantage of such a system includes the significant reduction in time required to change waveform parameters. This provides an easily tunable source of high-fidelity stable voltage waveforms at low temperature for future applications including validation of components for use in quantum technologies.
In this paper we present the uncertainty evaluation of a new cryogenic current comparator (CCC) bridge designed to compare two-terminal 1 MOmega and 10 MOmega standard resistors with the quantized Hall resistance (QHR) and then scale from these values to other values between 10 kOmega and 1 GOmega.
11 August 2023 in IEEE Transactions on Instrumentation and Measurement
This paper presents an improved system for traceable calibrations of weld current monitors/testers at short duration high pulsed currents developed recently at the Instituto Nacional de Tecnología Industrial (INTI) in Argentina. The calibration system has an increased capability of generating short duration high pulsed currents of up to 25 kA, and an improved uncertainty of 0.2 % of the calibration system for the measurement of their root-mean-square values. The performance evaluation of the improved calibration system conducted at INTI and also at the National Research Council (NRC) Canada is presented. Its use is intended for providing a traceable and IEC/ISO 17025 Standard compliant calibration service to the automotive industry in Argentina.
Este artículo presenta el estudio exploratorio de la cobertura geográfica de una red de LoRaWAN ubicada en una zona urbana. Se obtuvo un alcance estable de hasta 19 km utilizando un Field Tester especialmente desarrollado para este trabajo y vinculado a plataformas on-line para el registro y visualización de la información. Se observó una dependencia entre la máxima distancia y la instalación del gateway.