We used absolute primary acoustic gas thermometry (AGT) to calibrate a Pt–Co resistance thermometer on the thermodynamic temperature scale by measuring the speed of sound in helium at a temperature T* chosen to be near the temperature of the triple point of neon, TNe. Prior to the present AGT, the Pt–Co thermometer was used with a neon triple-point cell as part of an interlaboratory comparison. Taken together, the results of the interlaboratory comparison and the present AGT redetermined the thermodynamic temperature TNe = (24.555 15 ± 0.000 24) K. This new value of TNe is consistent with other recent determinations obtained with various primary methods. After completing the AGT thermodynamic calibration, we used the Pt–Co thermometer to link T* to the temperature ratios measured by single-pressure refractive-index gas thermometry (SPRIGT) in a different laboratory. (Gao et al 2020 Metrologia57 065006) Now, the T*-linked SPRIGT system can calibrate other thermometers on the thermodynamic temperature scale T in the range 5 K ⩽ T ⩽ TNe without using the international temperature scale ITS-90. At most temperatures in this range, the uncertainties of the T*-linked SPRIGT system are smaller than those of the ITS-90 systems used by National Metrology Institutes to calibrate resistance thermometers.
We describe a primary gas pressure standard based on the measurement of the refractive index of helium gas using a microwave resonant cavity in the range between 500 Pa and 20 kPa. To operate in this range, the sensitivity of the microwave refractive gas manometer (MRGM) to low-pressure variations is substantially enhanced by a niobium coating of the resonator surface, which becomes superconducting at temperatures below 9 K, allowing one to achieve a frequency resolution of about 0.3 Hz at 5.2 GHz, corresponding to a pressure resolution below 3 mPa at 20 Pa. The determination of helium pressure requires precise thermometry but is favored by the remarkable accuracy achieved by ab initio calculations of the thermodynamic and electromagnetic properties of the gas. The overall standard uncertainty of the MRGM is estimated to be of the order of 0.04%, corresponding to 0.2 Pa at 500 and 8.1 Pa at 20 kPa, with major contributions from thermometry and the repeatability of microwave frequency measurements. A direct comparison of the pressures realized by the MRGM with the reference provided by a traceable quartz transducer shows relative pressure differences between 0.025% at 20 kPa and −1.4% at 500 Pa.
Vers un étalon quantique pour des mesures absolues de pression: La mesure de pression joue un rôle critique dans de nombreux domaines scientifiques et industriels. Un exemple récent concerne les mesures de la constante de Boltzmann réalisées pour la redéfinition du kelvin. Ces mesures nécessitent une connaissance de la température et de la pression au meilleur niveau d’exactitude. Dans l’industrie aéronautique, la pression fournit l’altitude et la vitesse. Dans l’industrie nucléaire, une excellente connaissance de la pression est indispensable pour garantir la sûreté de fonctionnement des centrales. Il existe ainsi une forte demande de l’industrie et des laboratoires de métrologie pour une amélioration des incertitudes d’étalonnage dans le domaine allant de 200 Pa à 2·104 Pa (de 0,002 atm à 0,2 atm). L’exactitude des mesures nécessite qu’elles soient traçables à un étalon primaire. Le principe du nouvel étalon est basé d’une part sur la mesure de l’indice de réfraction de l’hélium sous forme gazeuse, fonction de la pression et de la température et d’autre part sur les calculs ab initio des propriétés thermophysiques de l’hélium. Ces derniers permettent d’atteindre un niveau d’exactitude jusqu’ici inégalé. Le but de ce projet consiste à développer un nouveau banc d’étalonnage en basses pressions, basé sur les avancées métrologiques récentes réalisées par le Laboratoire commun de métrologie LNE-CNAM sur les cavités micro-ondes où la traçabilité des mesures sera liée aux étalons de fréquence, ce qui permettra de réduire de manière importante les incertitudes. Le système sera conçu pour être facilement transportable de manière à faciliter les comparaisons inter-laboratoire.