6 December 2007 in IEEE Transactions on Applied Superconductivity
A 75-ch SQUID biomagnetometer system for the measurement of the cervical spinal cord evoked magnetic field (SCEF) was developed for the purpose of the noninvasive functional diagnosis of the spinal cord. The sensor array has 25 SQUID vector sensors arranged along the cylindrical surface to fit to the shape of the subject's neck. The magnetic fields, not only in the direction radial to the subject's body surface but also in the tangential direction, are observed in the area of 80 mm times 90 mm at one time. The dewar has a unique shape with a cylindrical main body and a protrusion from its side surface. The sensor array is installed in the protruded part. This design is optimized to detect magnetic signals at the back of the neck of the subject sitting in a reclining position. We applied the developed SQUID system to the cervical SCEF measurement of normal subjects who were given electric pulse stimulation to their median nerves at the wrists. The evoked magnetic signals were successfully detected at the cervixes of all subjects. A characteristic pattern of transition of the SCEF distribution was observed as a reproducible result and the signal components propagating along the spinal cord were found in the time varying SCEF distribution. We expect that the investigation of the propagating signal components would help to establish a noninvasive functional diagnosis of the spinal cord.
2 February 2021 in IEEE Transactions on Applied Superconductivity
We developed a superconducting quantum interference device (SQUID)-based biomagnetometer system for magnetoneurograms. The system was improved from a previously reported prototype magnetospinograph (MSG)/neurograph (MNG) system to satisfy the additional requirements that arose when we investigated practical clinical applications of the MSG/MNG in a hospital. The dimensions of the vector-type SQUID flux sensor array were increased to 150 mm × 188 mm, and the shape of the cryostat was modified so that the peripheral nerves at different body parts of a subject who assumed various postures, including positions provoking neurological symptoms, could be examined. In this paper, the improved MNG system is described in detail. Functional imaging of the brachial plexuses of subjects whose arms were in the abduction and external rotation position, which is one of the typical symptom-provocative positions for the shoulder, was achieved using the improved MNG system.
14 February 2022 in Superconductor Science and Technology
An important parameter of Josephson junctions (JJs) is the product of normal state resistance (R n) and critical current (I c) for designing superconductor analogue devices or digital circuits. Determination of R n and I c from voltage–current (U–I) characteristic curves often faces difficulties; in particular I c is considerably reduced by intrinsic thermal or extrinsic electrical noises. Here, we propose a standard measurement method of R n and intrinsic critical current (I ci) for high-T c superconductor (HTS) grain boundary JJs operated in liquid nitrogen and low-T c superconductor (LTS) multilayer superconductor/normal-conductor/superconductor (SNS) JJs in liquid helium. The applicable condition of this method is that both HTS and LTS JJs have U–I curves compatible with resistively-shunted junction (RSJ) model. Both R n and I ci values are extracted by combining a geometric mean criterion to select a data set and a least-squares fitting method with the RSJ model, eliminating two distortion effects on U–I curves: noise-rounding and self-heating. The combined method ensures relative standard uncertainty values of 1.9% for R n and 8% for I ci or better, when the users follow the standard protocol. It is demonstrated that the combined method is valid for d-wave HTS JJs near 77 K, regardless of peculiarities such as 0–π junction transition in d-wave superconductors at lower temperatures, and s-wave LTS SNS JJs with a wide range of junction parameters. This work is the first step to facilitate standardization for superconductor electronics with JJs.
10 April 2023 in IEEE Transactions on Instrumentation and Measurement
We proposed a spherical coil array for determining the positions, orientations, and sensitivities of the magnetometers for whole-head magnetoencephalograph (MEG) systems. The coil array comprises 16 concentric and symmetrically oriented circular coils with a diameter of 150 mm, which is larger than those of conventional calibration coil arrays, such as 5 mm or 60 mm. The accuracy of the calibration was expected to improve because the relative mechanical errors in machining and assembly could be reduced when the coil diameter was increased. Moreover, we also proposed a method to estimate the uncertainties in calibration, and it was demonstrated that the sensor parameters of each sensor were successfully obtained with uncertainties using a spherical calibration coil array. Additionally, the accuracy of the calibration was evaluated using a dry-type MEG phantom, and as expected, it was found that the accuracy was improved from that of a conventional calibration coil array composed of an assembly of multiple 3-axis bobbins.
9 June 2021 in IEEE Transactions on Magnetics
Magnetoencephalography (MEG) is highly sensitive to even a small amount of magnetic noise, and thus, a high-performance mag-netically shielded room (MSR) is essential. The shielding effect of a novel MSR should be predicted to verify whether it satisfies the required specifications of the MEG at the design stage. Therefore, an accurate prediction method based on the properties of shielding materials and the shape of the MSR is necessary. In this paper, we proposed a prediction method using a simple formula and permeability measured against very small magnetic field fluctuations. Using a ring-shaped test piece, permeability was measured against a very small magnetic field corresponding to the Rayleigh region, and the initial permeability and Rayleigh constant were estimated to determine the permeability calculated by the formula. The magnetic field generated inside the material was estimated by magnetic analysis, and it was confirmed that the permeability of the Rayleigh region should be used in the formula. To verify the prediction accuracy, we examined a simple example of a cylindrical shell by comparing the calculated results with the experimental results. It was confirmed that the pro-posed method is more accurate than the conventional method using the value of μ at 4 mA/cm.
10 August 2020 in IEEE Transactions on Magnetics
Mu-metal is used to shield highly sensitive machines against magnetic noises. Magnetoencephalography is an application that uses mu-metal to construct a magnetically shielded room (MSR). Each layer of multi-layer MSR is exposed to a different strength of magnetic noise. The outer layers are exposed to higher noise levels, whereas the inner layers are exposed to lower noise levels. The mu-metal in each layer of the MSR is required to have different magnetic properties depending on the magnetic noise level to which it is exposed. In this study, we propose the use of the index of permeability μ under a very-low-frequency micromagnetic field to evaluate the magnetic properties of mu-metals used in the innermost layer of the MSR. A measurement system for μ is outlined in this paper. Measurements were conducted using different heat treatments. Three test pieces of mu-metal from the same lot were heat treated in a hydrogen atmosphere and subjected to different isothermal conditions and cooling rates. The measurement results demonstrated that the test pieces exhibited different magnetic properties at very slow and micromagnetic fields, and the properties of the mu-metal could be controlled by heat treatment.