Positron emission tomography (PET) is a well established imaging technique for range monitoring in hadrontherapy. Multiple fields are standard protocol in treatments, but because of washout and residual activity background from previous irradiation plans, to this date quantitative verification of the particle range for each beam field is still an open issue. In this paper, a new method for the evaluation with a PET detector of the activity range of the second field of a treatment is discussed. Two treatment plans with two parallel-opposed beam fields, one using protons and one using carbon ions, were delivered on PMMA phantoms. In both cases, the second beam field was extracted from the first irradiation residual activity and compared with a reference image, obtained from the experimental acquisition of the second field alone. Results demonstrate good agreement between the extracted second field and the reference image, with average difference in the activity range along the preferential direction of the beam less than 0.5 mm for protons, and 1.5 mm for carbon ions. Without taking into account any preferential direction, differences within 0.5 mm were found for both cases. The method will soon be tested with nonhomogeneous phantoms and, subsequently, with in-vivo clinical data.
In a carbon ion treatment the nuclear fragmentation of both target and beam projectiles impacts on the dose released on the tumour and on the surrounding healthy tissues. Carbon ion fragmentation occurring inside the patient body has to be studied in order to take into account this contribution. These data are also important for the development of the range monitoring techniques with charged particles. The production of charged fragments generated by carbon ion beams of 115-353 MeV/u kinetic energy impinging on carbon, oxygen and hydrogen targets has been measured at the CNAO particle therapy center (Pavia, Italy). The use of thin targets of graphite (C), PMMA (C2O5H8) and polyvinyl-toluene (plastic scintillator, CbHa) allowed to measure fragments production cross sections, exploiting a Time of Flight (ToF) technique. Plastic scintillator detectors have been used to perform the ToF measurements, while LYSO crystals have been used for the deposited energy measurement and to perform particle identification. Cross sections have been measured at 90 and 60 degrees with respect to the beam direction. The measured proton, deuteron and triton differential production cross sections on C, O and H, obtained exploiting the target subtraction strategy, are presented here as a function of the fragment kinetic energy.
In-beam PET monitoring can provide early treatment assessment in proton therapy. However, open ring configurations and low positron emitter production yields degrade image quality, hampering the assessment accuracy. To achieve the highest precision for range monitoring, it is compulsory to mitigate image noise and compensate for data truncation. The goal of this work is to study the performance of state-of-the-art algorithms for in-beam PET image reconstruction, by evaluating the impact of the system response model and assessing the accuracy of range measurements. The approaches investigated here were Maximum-A-Posteriori algorithms combined with Total-Variation and Median-Root Priors. Maximum-Likelihood-Expectation-Maximization was used as reference. To compute the system matrix, different models were compared: a precise Monte Carlo model, and Single-Ray Tracing with and without Gaussian blurring in image space. The proposed methods were tested on simulations of Spread-Out-Bragg-Peaks delivered on phantoms. The I3PET scanner geometry was used as a case study. After image post-processing, the explored methods delivered similar results. Our work demonstrates the feasibility and reliability of using a simple and fast reconstruction method to perform range evaluations, given the correct image post-processing.