This work describes a compact and low-power current feedback instrumentation amplifier (CFIA) that can effectively eliminate differential electrode offset (DEO), amplifier input offset, and light motion artifacts. The proposed approach is using a local feedback loop in the first stage of the CFIA which can eliminate DC and low frequency offset whereas by-passes useful components such as EEG and ECG signals. As a result, the DEO and amplifier input offset can be eliminated in the first stage of the CFIA while still amplifying the desired signal components. This enables offset cancellation operation with low bias current that leads to higher energy efficiency compared to existing CFIA offset cancellation schemes. The proposed CFIA is implemented using CMOS 0.18Jm technology, that can eliminated up to 250mV DEO with supply voltage of 1.8V. Measurement results show CMRR 120 dB (without DEO) and 95 dB (with DEO of 250mV) at 10 Hz, input noise 3.8 JVrms, power consumption 12.7, and NEF 39 with core area of 73.6 10-3 mm2. Furthermore, the actual EEG and ECG waveforms are obtained using the proposed CFIA under an actual clinical setting.
24 April 2015 in Analog Integrated Circuits and Signal Processing