Press-fitted implants are implanted by impaction to ensure adequate seating, but without overloading the components, the surgeon, or the patient. To understand this interrelationship a uniaxial discretised model of the hammer/introducer/implant/bone/soft-tissues was developed. A parametric analysis of applied energy, component materials and geometry, and interactions between implant and bone and between bone and soft-tissues was performed, with implant seating and component stresses as outcome variables. To reduce the impaction effort (energy) required by the surgeon for implant seating and also reduce stresses in the hardware the following outcomes were observed: Reduce energy per hit with more hits / Increase hammer mass / Decrease introducer mass / Increase implant-bone resistance (eg stem roughness). Hardware stiffness and patient mechanics were found to be less important and soft tissue forces, due to inertial protection by the bone mass, were so low that their damage would be unlikely. This simple model provides a basic understanding of how stress waves travel through the impacted system, and an understanding of their relevance to implantation technique and component design.
Implants anchored by press-fit are predominantly implanted by impaction. This method allows sufficiently high forces to be generated easily by the surgeon. Suitable impaction should provide adequate implant seating without damaging the patient (tissues), the implant and implantation system, or the surgeon. However, issues have been documented for all of these factors. In this study a model to predict implant seating is developed, given an applied impaction impulse, the mass of the accelerated components and the push-in resistance force. The model was validated against experimental data for a contemporary femoral stem implanted in a polyurethane foam surrogate for bone, with the input parameters varied. The model tended to overestimate seating but represented seating patterns well. The model can be used to estimate implant seating using easily measured parameters and could be useful in the design of implantation systems, and in optimising impaction strategies.
Background: Considerable variation exists in the proximal and distal fixture designs used in biomechanical testing of diaphyseal femoral implants. In most of them a load is applied to the femoral head, but muscle forces are rarely represented, so bending of the diaphysis is unphysiologically high. Methods: This study investigated twelve testing configurations reported in the literature using finite element and analytical models of the femoral diaphysis and a lateral fixation plate. The models differed in proximal and distal displacement constraints, femoral orientation, and load application. Internal force distributions, as well as bone and implant stresses and bone strains, were calculated for each configuration. Results: Common experimental conditions, in which the load is applied to the femoral head, generally produced high bending moments, resulting in, what we interpret to be, unrealistically high bone and implant stresses and bone strains. In contrast, axial loading along the femoral shaft generated substantially lower bending, which we propose to be more physiological, particularly in two such cases with reduced rotational constraint. Conclusions: Their constant bending distribution should improve reproducibility and their implementation should provide a practical approach to biomechanical testing of diaphyseal femoral implants.