The current pre-clinical testing standards for total hip replacements (THRs), ISO standards, use simplified loading waveforms that do not fully replicate real-world biomechanics. These standards provide a benchmark of data that may not accurately predict in vivo wear, necessitating the evaluation of physiologically relevant loading conditions. Previous studies have incorporated activities of daily living (ADLs) such as walking, jogging and stair negotiation into wear simulations. However, these studies primarily used simplified adaptations that increased axial forces and applied accelerated sinusoidal waveforms, rather than fully replicating the complex kinematics experienced by THR patients. To address this gap, this study applied patient-derived ADL profiles—jogging and stair negotiation—using a three-station hip simulator, obtained through 3D motion analysis of total hip arthroplasty patients, processed via a musculoskeletal multibody modelling approach to derive realistic hip contact forces (HCFs). The results indicate that jogging significantly increased wear rates compared to the ISO walking gait waveform, with wear increasing from 15.24 ± 0.55 to 28.68 ± 0.87 mm3/Mc. Additionally, wear was highly sensitive to changes in lubricant protein concentration, with an increase from 17 g/L to 30 g/L reducing wear by over 60%. Contrary to predictive models, stair descent resulted in higher volumetric wear (8.62 ± 0.43 mm3/0.5 Mc) compared to stair ascent (4.15 ± 0.31 mm3/0.5 Mc), despite both profiles having similar peak torques. These findings underscore the limitations of current ISO standards in replicating physiologically relevant wear patterns. The application of patient-specific loading profiles highlights the need to integrate ADLs into pre-clinical testing protocols, ensuring a more accurate assessment of implant performance and longevity.
Morse-type tapers at the head–stem junction in total hip replacements (THRs) provide many benefits to permit a successful surgical outcome. However, with the introduction of modular tapered devices comes complications associated with fluid ingress and motion at the interface that can cause fretting corrosion, which has been implicated in clinical failure. Increased surface roughness amplitude (Ra) and angular mismatch to ensure taper contact closer to the equator of the femoral head are design features introduced for use with ceramic heads but have been adopted by metal head couples. While increased surface roughness amplitude has been found to contribute to fretting corrosion, there is a distinct lack of systematic studies investigating the interactions between angular mismatch and Ra. This study measured the fretting corrosion and motion response of clinically representative samples, in part reference to ASTM F1875, when subjected to uniaxial incremental dynamic loading. The fretting corrosion response was measured in situ with an integrated three electrode electrochemical cell. Motion at the head–neck interface was measured with a bespoke motion measurement solution based on eddy-current principles which uses four sensors to allow motion to be fully characterised in three dimensions. Key findings from this study included a 5–10-fold increase in current measured in the increased roughness amplitude samples, suggesting an increased susceptibility to fretting corrosion without a corresponding increase in motion. The distal samples engaged around the opening of the taper interface and presented the lowest current measurements but most off-axis subsidence. Findings from this study indicate that optimisation of the taper interfaces in THR, in terms of fretting corrosion and motion, can be made and can be assessed using short-term preclinical tests.