Objectives The main objective of the study was to compare the accuracy of full-arch digital implant impressions for fixed dental prosthesis under in vitro and in vivo conditions. Material and Methods Eight patients (5 women and 3 men) with at least one edentulous arch and with 4-6 osseointegrated implants participated in this study. For each edentulous arch (n=10) experimental screw-retained titanium bar with attached 4 scan bodies was fabricated. The bar containing 4 scan bodies was screw-retained intraorally on implants and scanned with Trios 3 intraoral scanner 8 times (IOS group, in vivo). Then the bar was attached to the master cast and scanned 8 times again with the same intraoral scanner (MIOS group, in vitro). Finally, the bar with scan bodies was scanned 8 times with a laboratory scanner (reference). Precision and trueness were calculated for 3 distances and 3 angles between the scan bodies (1-2, 1-3, and 1-4) in IOS and MIOS groups. Results Precision and trueness for the largest distance (1-4) were found to be 44±18 µm and 32±19 µm for the IOS group, and 31±16 µm and 30±14µm for MIOS group, respectively. Precision and trueness for the angle between the most distant scan bodies (1-4) were 0.22±0.14° and 0.18±0.10° for the IOS group, and 0.16±0.11° and 0.07±0.05° for MIOS group, respectively. Conclusions Intraoral conditions moderately affected the precision and trueness of Trios 3 (3Shape) intraoral scanner. Results of in vitro accuracy studies cannot be directly transferred to the clinical field.
1 January 2019 in The Journal of Advanced Prosthodontics